Semi‐automated vertex placement for lattice radiotherapy and dosimetric verification using 3D polymer gel dosimetry
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14489 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Tenzin Kunkyab und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Computer Science, Mathematics, Physics and Statistics, The University of British Columbia Okanagan, Kelowna, British Columbia, Canada. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiotherapy, Intensity-Modulated, Polymers. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: BC Cancer Kelowna, Moss Rock Park Foundation.
Abstract (Originalsprache)
Purpose
To evaluate the feasibility of an open‐source, semi‐automated, and reproducible vertex placement tool to improve the efficiency of lattice radiotherapy (LRT) planning. We used polymer gel dosimetry with a Cone Beam CT (CBCT) readout to commission this LRT technique. Material and methods We generated a volumetric modulated arc therapy (VMAT)‐based LRT plan on a 2 L NIPAM polymer gel dosimeter using our Eclipse Acuros version 15.6 AcurosXB beam model, and also recalculated the plan with a pre‐clinical Acuros v18.0 dose calculation algorithm with the enhanced leaf modelling (ELM). With the assistance of the MAAS‐SFRThelper software, a lattice vertex diameter of 1.5 cm and center‐to‐center spacing of 3 cm were used to place the spheres in a hexagonal, closed packed structure. The verification plan included four gantry arcs with 15°, 345°, 75°, 105° collimator angles. The spheres were prescribed 20 Gy to 50% of their combined volume. The 6 MV Flattening Filter Free beam energy was used to deliver the verification plan. The dosimetric accuracy of the LRT delivery was evaluated with 1D dose profiles, 2D isodose maps, and a 3D global gamma analysis.
Results
Qualitative comparisons between the 1D dose profiles of the Eclipse plan and measured gel showed good consistency at the prescription dose mark. The average diameter measured 13.3 ± 0.2 mm (gel for v15.6), 12.6 mm (v15.6 plan), 13.1 ± 0.2 mm (gel for v18.0), and 12.3 mm (v18.0 plan). 3D gamma analysis showed that all gamma pass percent were > 95% except at 1% and 2% at the 1 mm distance to agreement criteria.
Conclusion
This study presents a novel application of gel dosimetry in verifying the dosimetric accuracy of LRT, achieving excellent 3D gamma results. The treatment planning was facilitated by publicly available software that automatically placed the vertices for consistency and efficiency.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Computer Science, Mathematics, Physics and Statistics, The University of British Columbia Okanagan, Kelowna, British Columbia, Canada.
- BC Cancer, Kelowna, British Columbia, Canada.
- Varian Medical Systems, Palo Alto, California, USA.
- Department of Surgery, The University of British Columbia, Vancouver, British Columbia, Canada.
Förderung
- BC Cancer Kelowna
- Moss Rock Park Foundation
Zitieren
APA
Kunkyab, T., Magliari, A., Jirasek, A., Mou, B., & Hyde, D. (2024). Semi‐automated vertex placement for lattice radiotherapy and dosimetric verification using 3D polymer gel dosimetry. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14489
Vancouver
Kunkyab T, Magliari A, Jirasek A, Mou B, Hyde D. Semi‐automated vertex placement for lattice radiotherapy and dosimetric verification using 3D polymer gel dosimetry. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14489
BibTeX
@article{Kunkyab2024,
author = {Kunkyab, Tenzin and Magliari, Anthony and Jirasek, Andrew and Mou, Benjamin and Hyde, Derek},
title = {Semi‐automated vertex placement for lattice radiotherapy and dosimetric verification using 3D polymer gel dosimetry},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14489}
}
EUVIMED