Beam commissioning of a new compact scanned proton therapy system with four different dose calculation algorithms
Journal of Applied Clinical Medical Physics · 2026-01 · 27 (1)
DOI: 10.1002/acm2.70433 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 19.12.2025. Verfasst von Ryo Tanokura und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Central Japan International Medical Center, Minokamo, Gifu, Japan. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Proton Therapy, Algorithms, Phantoms, Imaging. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The Central Japan International Medical Center (Gifu prefecture, Japan) as the 20th Proton Therapy Institution in Japan, has newly installed a compact scanned proton therapy system, ProBeam 360°™. We installed two commercial treatment planning systems (TPSs): Eclipse and RayStation. No commissioning results have been reported between Varian's newest compact proton therapy system and the two TPSs. Purpose We identified the optimal dose calculation algorithm for clinical use by evaluating the dose agreements between four dose calculation algorithms provided by two TPSs and the measured values using our new machine.
Methods
The two TPSs each had a pencil beam algorithm (PBA) and Monte Carlo (MC) dose engines (four in total). Validation plans were created for 26 box targets (with and without a 50‐mm range shifter) in a homogeneous phantom optimized using Eclipse. We created three realistic clinical plans consisting of 42 beams in mock patients with prostate, head and neck, and C‐shaped targets. We validated 12 heterogeneous plans for the lung and bone regions using all three optimizations and four dose algorithms with the TPSs. Measurements were performed on the lateral profiles at the isocenter plane. The calculated doses for each algorithm were compared with the measured doses.
Results
Each dose algorithm achieved an average gamma score of at least 95.6 ± 5.0% or higher (RayStation PBA) at 3%/3 mm in the box plans. For the patient and heterogeneous plans, the average gamma score exceeded 95% for only two MC algorithms. RayStation MC showed the highest dose calculation accuracy in both homogeneous and heterogeneous regions.
Conclusions
All calculated doses of the box plans were in good agreement with the measurements, except for the RayStation PBA. For clinical use, RayStation MC is the preferred option for both homogeneous and heterogeneous cases.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Technology, Central Japan International Medical Center, Minokamo, Gifu, Japan.
- Department of Radiation Oncology, Central Japan International Medical Center, Minokamo, Gifu, Japan.
- Department of Radiotherapy, Medical Co. Hakuhokai, Osaka Proton Therapy Clinic, Konohana-ku, Osaka, Japan.
- Medical Physics Laboratory, Division of Health Science, Graduate School of Medicine, The University of Osaka, Suita-Shi, Osaka, Japan.
- Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Sakae-machi, Fukushima, Japan.
ORCID-Profile: Yuki Tominaga
Zitieren
APA
Tanokura, R., Horita, M., Matsumoto, M., Miyahara, A., Furukawa, S., Ido, Y., Matsumoto, A., Ogawa, S., Makino, W., Matsui, Y., Tominaga, Y., Kato, T., & Fuwa, N. (2026). Beam commissioning of a new compact scanned proton therapy system with four different dose calculation algorithms. Journal of Applied Clinical Medical Physics, 27(1). https://doi.org/10.1002/acm2.70433
Vancouver
Tanokura R, Horita M, Matsumoto M, Miyahara A, Furukawa S, Ido Y, et al. Beam commissioning of a new compact scanned proton therapy system with four different dose calculation algorithms. Journal of Applied Clinical Medical Physics. 2026;27(1). doi:10.1002/acm2.70433
BibTeX
@article{Tanokura2026,
author = {Tanokura, Ryo and Horita, Masato and Matsumoto, Makoto and Miyahara, Akari and Furukawa, Shinji and Ido, Yasushi and Matsumoto, Akira and Ogawa, Shinichi and Makino, Wataru and Matsui, Yoshihito and Tominaga, Yuki and Kato, Takahiro and Fuwa, Nobukazu},
title = {Beam commissioning of a new compact scanned proton therapy system with four different dose calculation algorithms},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {1},
doi = {10.1002/acm2.70433}
}
EUVIMED