Dosimetric characterization of a new surface‐conforming electron MLC prototype
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14173 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Holly M Parenica Paschal und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, School of Medicine, The University of Texas Health Science Center at San Antonio, San Antonio, Texas, USA. Hauptthemen laut MEDLINE: Electrons, Particle Accelerators. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
The purpose is to reduce normal tissue radiation toxicity for electron therapy through the creation of a surface‐conforming electron multileaf collimator (SCEM). The SCEM combines the benefits of skin collimation, electron conformal radiotherapy, and modulated electron radiotherapy. An early concept for the SCEM was constructed. It consists of leaves that protrude towards the patient, allowing the leaves to conform closely to irregular patient surfaces. The leaves are made of acrylic to decrease bremsstrahlung, thereby decreasing the out‐of‐field dose. Water tank scans were performed with the SCEM in place for various field sizes for all available electron energies (6, 9, 12, and 15 MeV) with a 0.5 cm air gap to the water surface at 100 cm source‐to‐surface distance (SSD). These measurements were compared with Cerrobend cutouts with the field size‐matched at 100 and 110 cm SSD. Output factor measurements were taken in solid water for each energy at d max for both the cerrobend cutouts and SCEM at 100 cm SSD. Percent depth dose (PDD) curves for the SCEM shifted shallower for all energies and field sizes. The SCEM also produced a higher surface dose relative to Cerrobend cutouts, with the maximum being a 9.8% increase for the 3 cm × 9 cm field at 9 MeV. When compared to the Cerrobend cutouts at 110 cm SSD, the SCEM showed a significant decrease in the penumbra, particularly for lower energies (i.e., 6 and 9 MeV). The SCEM also showed reduced out‐of‐field dose and lower bremsstrahlung production than the Cerrobend cutouts. The SCEM provides significant improvement in the penumbra and out‐of‐field dose by allowing collimation close to the skin surface compared to Cerrobend cutouts. However, the added scatter from the SCEM increases shallow PDD values. Future work will focus on reducing this scatter while maintaining the penumbra and out‐of‐field benefits the SCEM has over conventional collimation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, School of Medicine, The University of Texas Health Science Center at San Antonio, San Antonio, Texas, USA.
ORCID-Profile: Holly M Parenica Paschal, Daniel Saenz
Förderung
- NCI NIH HHS
Zitieren
APA
Paschal, H. M. P., Kabat, C. N., Martin, T., Saenz, D., Myers, P., Rasmussen, K., Stathakis, S., Bonnen, M., Papanikolaou, N., & Kirby, N. (2024). Dosimetric characterization of a new surface‐conforming electron MLC prototype. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14173
Vancouver
Paschal HMP, Kabat CN, Martin T, Saenz D, Myers P, Rasmussen K, et al. Dosimetric characterization of a new surface‐conforming electron MLC prototype. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14173
BibTeX
@article{Paschal2024,
author = {Paschal, Holly M Parenica and Kabat, Christopher N and Martin, Thomas and Saenz, Daniel and Myers, Pamela and Rasmussen, Karl and Stathakis, Sotirios and Bonnen, Mark and Papanikolaou, Nikos and Kirby, Neil},
title = {Dosimetric characterization of a new surface‐conforming electron MLC prototype},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14173}
}
EUVIMED