Sensitivity and specificity of secondary dose calculation for head and neck treatment plans

Sonja Wegener iD, Ruaa Abu Rashed, Otto A Sauer, Gary Razinskas iD

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.14139 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Sonja Wegener und 3 weiteren Autor:innen. Beteiligte Einrichtung: University Hospital Wurzburg, Department of Radiation Oncology, Wuerzburg, Germany. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Secondary dose calculation (SDC) with an independent algorithm is one option to perform plan‐specific quality assurance (QA). While measurement‐based QA can potentially detect errors in plan delivery, the dose values are typically only compared to calculations on homogeneous phantom geometries instead of patient CT data. We analyzed the sensitivity and specificity of an SDC software by purposely introducing different errors and determined thresholds for optimal decisions.

Methods

Thirty head and neck VMAT plans and 30 modifications of those plans, including errors related to density and beam modelling, were recalculated using RadCalc with a Monte Carlo algorithm. Decision thresholds were obtained by receiver operating characteristics (ROC) analysis. For comparison, measurement‐based QA using the ArcCHECK phantom was carried out and evaluated in the same way.

Results

Despite optimized decision thresholds, none of the systems was able to reliably detect all errors. ArcCHECK analysis using a 2%/2 mm criterion with a threshold of 91.1% had an area under the curve (AUC) of 0.80. Evaluating differences in recalculated and planned DVH parameter of the target structures in RadCalc with a 2% threshold an AUC of 0.86 was achieved. Out‐of‐field deviations could be attributed to weaknesses in the beam model.

Conclusions

Secondary dose calculation with RadCalc is an alternative to established measurement‐based phantom QA. Different tools catch different errors; therefore, a combination of approaches should be preferred.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedHumansRadiotherapy Planning, Computer-AssistedQuality Assurance, Health CareSoftwareSensitivity and SpecificityRadiotherapy Dosage

Stichwörter der Autor:innen

ROC analysisplan-specific quality assurancesecondary dose calculation

Autor:innen und Institutionen

  • University Hospital Wurzburg, Department of Radiation Oncology, Wuerzburg, Germany.
  • Ernst-Abbe-Hochschule Jena, Jena, Germany.

ORCID-Profile: Sonja Wegener, Gary Razinskas

Zitieren

APA

Wegener, S., Abu Rashed, R., Sauer, O. A., & Razinskas, G. (2023). Sensitivity and specificity of secondary dose calculation for head and neck treatment plans. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14139

Vancouver

Wegener S, Abu Rashed R, Sauer OA, Razinskas G. Sensitivity and specificity of secondary dose calculation for head and neck treatment plans. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14139

BibTeX
@article{Wegener2023,
  author = {Wegener, Sonja and Abu Rashed, Ruaa and Sauer, Otto A and Razinskas, Gary},
  title = {Sensitivity and specificity of secondary dose calculation for head and neck treatment plans},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14139}
}
Kostenlos starten →