Patient‐specific geometrical distortion corrections of MRI images improve dosimetric planning accuracy of vestibular schwannoma treated with gamma knife stereotactic radiosurgery
Journal of Applied Clinical Medical Physics · 2023
DOI: 10.1002/acm2.14072 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Mojtaba Safari und 3 weiteren Autor:innen. Beteiligte Einrichtung: Département de physique, de génie physique et d'optique, et Centre de recherche sur le cancer, Université Laval, Québec, Québec, Canada. Hauptthemen laut MEDLINE: Radiosurgery, Neuroma, Acoustic. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Science and Engineering Research Council of Canada.
Abstract (Originalsprache)
Purpose
To investigate the impact of MRI patient‐specific geometrical distortion (PSD) on the quality of Gamma Knife stereotactic radiosurgery (GK‐SRS) plans of the vestibular schwannoma (VS) tumors. Methods and materials Three open access datasets including the MPI‐Leipzig Mind‐Brain‐Body (318 patients), the slow event‐related fMRI designs dataset (62 patients), and the VS dataset (242 patients) were used. We used first two datasets to train a 3D convolution network to predict the distortion map of third dataset that were then used to calculate and correct the PSD. GK‐SRS plans of VS dataset were used to evaluate dose distribution of PSD‐corrected MRI images. GK‐SRS prescription dose of VS cases was 12 Gy. Geometric and dosimetric discrepancies were assessed between the dose distributions and contours before and after the PSD corrections. Geometry indices were center of the contours, Dice coefficient (DC), Hausdorff distance (HD), and dosimetric indices were , , , and doses, target coverage (TC), Paddick's conformity index (PCI), Paddick's gradient index (GI), and homogeneity index (HI).
Results
Geometric distortions of about 1.2 mm were observed at the air‐tissue interfaces at the air canal and nasal cavity borders. Average center of the targets was significantly distorted along the frequency encoding direction after the PSD‐correction. Average DC and HD metrics were 0.90 and 2.13 mm. Average , and in Gy significantly increased after PSD correction from 16.85 to 17.25, 12.30 to 12.77, and from 8.98 to 9.92. did not significantly change after the correction. Average TC and PCI significantly increased from 0.97 to 0.98, and 0.94 to 0.96. Average GI decreased significantly from 2.24 to 2.15 after PSD correction. However, HI did not significantly change after the correction.
Conclusion
The proposed method could predict and correct the PSD that indicates the importance of PSD correction before GK‐SRS plans of the VS patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Département de physique, de génie physique et d'optique, et Centre de recherche sur le cancer, Université Laval, Québec, Québec, Canada.
- Service de physique médicale et de radioprotection, Centre Intégré de Cancérologie, CHU de Québec-Université Laval et Centre de recherche du CHU de Québec, Québec, Québec, Canada.
- Department of Physics, Jackson State University, Mississippi, USA.
- Merit Health Central, Department of Radiation Oncology, Gamma Knife Center, Mississippi, USA.
- Clinical Research Development Unit of Tabriz Valiasr Hospital, Tabriz University of Medical Science, Tabriz, Iran.
ORCID-Profile: Mojtaba Safari
Förderung
- Natural Science and Engineering Research Council of Canada
Zitieren
APA
Safari, M., Fatemi, A., Afkham, Y., & Archambault, L. (2023). Patient‐specific geometrical distortion corrections of MRI images improve dosimetric planning accuracy of vestibular schwannoma treated with gamma knife stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14072
Vancouver
Safari M, Fatemi A, Afkham Y, Archambault L. Patient‐specific geometrical distortion corrections of MRI images improve dosimetric planning accuracy of vestibular schwannoma treated with gamma knife stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14072
BibTeX
@article{Safari2023,
author = {Safari, Mojtaba and Fatemi, Ali and Afkham, Younes and Archambault, Louis},
title = {Patient‐specific geometrical distortion corrections of MRI images improve dosimetric planning accuracy of vestibular schwannoma treated with gamma knife stereotactic radiosurgery},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
doi = {10.1002/acm2.14072}
}
EUVIMED