Comparative Evaluation of Four Stereotactic Radiosurgery Planning Approaches for Treatment of Posterior Choroidal Melanoma
Journal of Medical Radiation Sciences · 2026
DOI: 10.1002/jmrs.70076 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Medical Radiation Sciences, 2026. Verfasst von Robert Nigro und 5 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Therapy Service, Peter MacCallum Cancer Centre, Melbourne, Victoria, Australia. Hauptthemen laut MEDLINE: Uveal Melanoma, Radiotherapy Planning, Computer-Assisted, Radiosurgery, Organs at Risk. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
This was a comparative treatment planning study for posterior choroidal melanoma (CM) using four stereotactic approaches: Brainlab iPlan dynamic conformal arc therapy (DCAT), Brainlab Elements modulated arcs (BE), Varian Eclipse HyperArc modulated arcs (HA) and Gamma Knife ICON Leksell GammaPlan (GK). The clinical goal was to reduce dose to normal brain whilst maintaining other treatment planning goals.
Methods
Thirteen CM cases over 3 years (2015–2018) treated with DCAT prescribed 50 Gy in 5 fractions were selected. The DCAT technique involved a forward planning method whereas semi‐automated methods were used in BE, HA and GK. Target coverage and dose to critical organs were evaluated and compared across all approaches with the same set of structures used for each.
Results
Volume of brain receiving 10 Gy was DCAT 0.93 cc and 0.00 cc for GK, BE, HA ( p < 0.001). Brain D near−max GK 5.52 Gy, HA 6.41 Gy, BE 7.28 Gy, DCAT 13.73 Gy ( p = 0.001). The median Gradient Index (GI) was GK 2.79, DCAT 6.39, BE 5.07, HA 6.00 ( p < 0.001) with GK significantly different from each of the other approaches for PTV parameters. Further statistical significance was recorded for the lacrimal gland, contralateral lens and ipsilateral optic nerves.
Conclusions
Modulation tools in HA and BE enabled clinically meaningful reduction in dose to brain compared with DCAT whilst maintaining dose to PTV. GK produced highest PTV coverage with superior gradient index, however current methods for eye immobilisation and treatment delivery only allow for single fraction treatment with GK.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Radiation Therapy Service, Peter MacCallum Cancer Centre, Melbourne, Victoria, Australia.
- Sir Peter MacCallum Department of Oncology, The University of Melbourne, Melbourne, Victoria, Australia.
- Department of Physical Sciences, Peter MacCallum Cancer Centre, Melbourne, Victoria, Australia.
- RMIT University, School of Applied Sciences, Melbourne, Victoria, Australia.
- Division of Radiation Oncology, Peter MacCallum Cancer Centre, Melbourne, Victoria, Australia.
ORCID-Profile: Robert Nigro, Adam U Yeo
Zitieren
APA
Nigro, R., Thompson, K., Markham, C., Yeo, A. U., Sapkaroski, D., & Phillips, C. (2026). Comparative Evaluation of Four Stereotactic Radiosurgery Planning Approaches for Treatment of Posterior Choroidal Melanoma. Journal of Medical Radiation Sciences. https://doi.org/10.1002/jmrs.70076
Vancouver
Nigro R, Thompson K, Markham C, Yeo AU, Sapkaroski D, Phillips C. Comparative Evaluation of Four Stereotactic Radiosurgery Planning Approaches for Treatment of Posterior Choroidal Melanoma. Journal of Medical Radiation Sciences. 2026. doi:10.1002/jmrs.70076
BibTeX
@article{Nigro2026,
author = {Nigro, Robert and Thompson, Kenton and Markham, Cathy and Yeo, Adam U and Sapkaroski, Daniel and Phillips, Claire},
title = {Comparative Evaluation of Four Stereotactic Radiosurgery Planning Approaches for Treatment of Posterior Choroidal Melanoma},
journal = {Journal of Medical Radiation Sciences},
year = {2026},
doi = {10.1002/jmrs.70076}
}
EUVIMED