Commissioning and clinical evaluation of the IDENTIFY TM surface imaging system for frameless stereotactic radiosurgery
Journal of Applied Clinical Medical Physics · 2023-10 · 24 (10)
DOI: 10.1002/acm2.14058 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 08.06.2023. Verfasst von Elizabeth L Covington und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama, USA. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated, Brain Neoplasms. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To commission and assess the clinical performance of a new commercial surface imaging (SI) system by analyzing intra‐fraction motion from the initial cohort of patients treated with frameless stereotactic radiosurgery (fSRS).
Methods
The IDENTIFY TM SI system was commissioned for clinical use on an Edge (Varian Medical Systems, Palo Alto, CA) linear accelerator. All patients who received intracranial radiotherapy with HyperArc TM (Varian Medical Systems, Palo Alto, CA) were immobilized with the Encompass TM (Qfix, Avondale, PA) thermoplastic mask and monitored for intra‐fraction motion with SI. IDENTIFY TM log files were correlated with trajectory log files to correlate treatment parameters with SI‐reported offsets. IDENTIFY TM reported offsets were correlated with gantry and couch angles to assess system performance for obstructed and clear camera field of view. Data were stratified by race to evaluate performance differences due to skin tone.
Results
All commissioning data were found to meet recommended tolerances. IDENTIFY TM was used to monitor intra‐fraction motion on 1164 fractions from 386 patients. The median magnitude of translational SI reported offsets at the end of treatment was 0.27 mm. SI reported offsets were shown to increase when camera pods are blocked by the gantry with larger increases seen at non‐zero couch angles. With camera obstruction, the median magnitude of the SI reported offset was 0.50 and 0.80 mm for White and Black patients, respectively.
Conclusions
IDENTIFY TM performance during fSRS is comparable to other commercially available SI systems where offsets are shown to increase at non‐zero couch angles and during camera pod blockage.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama, USA.
- Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan, USA.
- Department of Neurosurgery, University of Alabama at Birmingham, Birmingham, Alabama, USA.
ORCID-Profile: Kristen O Riley, John B Fiveash
Zitieren
APA
Covington, E. L., Stanley, D. N., Sullivan, R. J., Riley, K. O., Fiveash, J. B., & Popple, R. A. (2023). Commissioning and clinical evaluation of the IDENTIFY TM surface imaging system for frameless stereotactic radiosurgery. Journal of Applied Clinical Medical Physics, 24(10). https://doi.org/10.1002/acm2.14058
Vancouver
Covington EL, Stanley DN, Sullivan RJ, Riley KO, Fiveash JB, Popple RA. Commissioning and clinical evaluation of the IDENTIFY TM surface imaging system for frameless stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. 2023;24(10). doi:10.1002/acm2.14058
BibTeX
@article{Covington2023,
author = {Covington, Elizabeth L and Stanley, Dennis N and Sullivan, Rodney J and Riley, Kristen O and Fiveash, John B and Popple, Richard A},
title = {Commissioning and clinical evaluation of the IDENTIFY TM surface imaging system for frameless stereotactic radiosurgery},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
volume = {24},
number = {10},
doi = {10.1002/acm2.14058}
}
EUVIMED