Implementation of superficial radiation therapy (SRT) using SRT‐100 Vision™ for non‐melanoma skin cancer in a Radiation Oncology clinic
Journal of Applied Clinical Medical Physics · 2023-06 · 24 (6)
DOI: 10.1002/acm2.13926 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 17.02.2023. Verfasst von Yongsook C Lee und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida, USA. Hauptthemen laut MEDLINE: Radiation Oncology, Skin Neoplasms. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This article describes our experience in implementation of superficial radiation therapy (SRT) using SRT‐100 Vision™ for non‐melanoma skin cancer.
Methods
Following the American Association of Physicists in Medicine Task Group‐61 protocol, absolute output (absorbed dose to water at surface (cGy/min)) was measured for three energies (50, 70, and 100 kV) and for six applicators (1.5–5.0 cm in diameter). Percent depth dose (PDD) and profiles were also measured. Timer testing and ultrasound testing were performed. A treatment time calculation worksheet was created. Quality assurance (QA) of SRT‐100 Vision was implemented. After treatment workflow for our clinic was developed, end‐to‐end (E2E) testing was performed using a Rando phantom. Considerations for treatment using SRT‐100 Vision were made.
Results
Absolute output (cGy/min) decreases as energy increases and applicator size decreases. Due to scatter from the applicator, PDD at depths ≤5 mm does not follow conventional trends but PDD at depths ≥15 mm increases with increasing applicator size. Profiles for the 5 cm applicator do not have strong dependence on depth except profiles at 5 mm for 50 kV. Timer/end errors are negligible for all three energies. Ultrasound images confirm allowed field of view and depth as well as no image artifacts and spatial integrity. Daily, monthly and annual QA of SRT‐100 Vision implemented in our clinic is listed in a table format. E2E testing results (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida, USA.
- Department of Radiation Oncology, Herbert Wertheim College of Medicine, Florida International University, Miami, Florida, USA.
Zitieren
APA
Lee, Y. C., Davis, S. D., Romaguera, W., Chaswal, V., Tolakanahalli, R., Gutierrez, A. N., & Kalman, N. S. (2023). Implementation of superficial radiation therapy (SRT) using SRT‐100 Vision™ for non‐melanoma skin cancer in a Radiation Oncology clinic. Journal of Applied Clinical Medical Physics, 24(6). https://doi.org/10.1002/acm2.13926
Vancouver
Lee YC, Davis SD, Romaguera W, Chaswal V, Tolakanahalli R, Gutierrez AN, et al. Implementation of superficial radiation therapy (SRT) using SRT‐100 Vision™ for non‐melanoma skin cancer in a Radiation Oncology clinic. Journal of Applied Clinical Medical Physics. 2023;24(6). doi:10.1002/acm2.13926
BibTeX
@article{Lee2023,
author = {Lee, Yongsook C and Davis, Stephen D and Romaguera, William and Chaswal, Vibha and Tolakanahalli, Ranjini and Gutierrez, Alonso N and Kalman, Noah S},
title = {Implementation of superficial radiation therapy (SRT) using SRT‐100 Vision™ for non‐melanoma skin cancer in a Radiation Oncology clinic},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
volume = {24},
number = {6},
doi = {10.1002/acm2.13926}
}
EUVIMED