Machine stability and dosimetry for ultra‐high dose rate FLASH radiotherapy human clinical protocol

Patrik Gonçalves Jorge iD, Reiner Geyer, Rémy Kinj, Luis Schiappacasse, Wendy Jeanneret-Sozzi, Jean Bourhis, Fernanda Herrera, François Bochud, Claude Bailat, Raphaël Moeckli

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70102 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Patrik Gonçalves Jorge und 9 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiation Physics, Lausanne University Hospital and Lausanne University, Lausanne, Switzerland. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiometry, Radiotherapy, Intensity-Modulated, Skin Neoplasms. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: ISREC Foundation, Biltema donation, Fondation pour le soutien de la recherche et du développement de l'oncologie (FSRDO).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The FLASH effect, induced by ultra‐high dose rate (UHDR) irradiations, offers the potential to spare normal tissue while effectively treating tumors. It is important to achieve precise and accurate dose delivery and to establish reliable detector systems, particularly for clinical trials needed to help the clinical transfer of FLASH‐Radiotherapy (FLASH‐RT). However, the use of monitoring chambers with UHDR beams is presently limited, leading to the reliance on passive dosimetry and machine stability.

Purpose

This study aimed to investigate the energy and output stability of a UHDR Mobetron (IntraOp, USA) and to compare it with its conventional dose rate (CDR) mode. Furthermore, we assessed the dosimetric accuracy of a human clinical protocol for FLASH‐RT.

Methods

Over a 26‐month duration, we assessed the short‐ and long‐term stability of the output and energy of the Mobetron system. Daily checks were conducted for 9 MeV CDR and UHDR. In parallel, the IMPulse clinical trial involving patients with skin metastases from melanoma was initiated. Prescription doses ranging from 22 to 28 Gy were administered. Pre‐, post‐, and in vivo dosimetry using alanine and thermoluminescent dosimeters (TLDs) was performed and compared to the prescription doses.

Results

Short‐term output fluctuations remained below 0.6 % and 1 % for 9 MeV CDR and UHDR, respectively. Long‐term output fluctuations were within 2 % and the long‐term energy fluctuations were below 2 mm (R 50 ) for both modes. The delivered doses of the IMPulse trial showed deviations below 4 % compared to prescription doses for all patients.

Conclusions

The Mobetron system demonstrated favorable short‐ and long‐term stability. There was a good agreement between the prescribed and the measured dose for the clinical IMPulse trial. The stability of this UHDR machine allows us to effectively conduct human clinical protocols as well as preclinical experiments, even in the absence of a real‐time monitoring system.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiometryRadiotherapy, Intensity-ModulatedSkin NeoplasmsMelanomaPhantoms, ImagingHumansRadiotherapy DosageClinical Protocols

Stichwörter der Autor:innen

FLASHUHDRdosimetrymonitoringreproducibilitystabilityultra‐high dose rate

Autor:innen und Institutionen

  • Institute of Radiation Physics, Lausanne University Hospital and Lausanne University, Lausanne, Switzerland.
  • Department of Radiation Oncology, Lausanne University Hospital and Lausanne University, Lausanne, Switzerland.

ORCID-Profile: Patrik Gonçalves Jorge

Förderung

  • ISREC Foundation
  • Biltema donation
  • Fondation pour le soutien de la recherche et du développement de l'oncologie (FSRDO)
  • IntraOp Medical Corporation

Zitieren

APA

Gonçalves Jorge, P., Geyer, R., Kinj, R., Schiappacasse, L., Jeanneret-Sozzi, W., Bourhis, J., Herrera, F., Bochud, F., Bailat, C., & Moeckli, R. (2025). Machine stability and dosimetry for ultra‐high dose rate FLASH radiotherapy human clinical protocol. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70102

Vancouver

Gonçalves Jorge P, Geyer R, Kinj R, Schiappacasse L, Jeanneret-Sozzi W, Bourhis J, et al. Machine stability and dosimetry for ultra‐high dose rate FLASH radiotherapy human clinical protocol. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70102

BibTeX
@article{GonalvesJorge2025,
  author = {Gonçalves Jorge, Patrik and Geyer, Reiner and Kinj, Rémy and Schiappacasse, Luis and Jeanneret-Sozzi, Wendy and Bourhis, Jean and Herrera, Fernanda and Bochud, François and Bailat, Claude and Moeckli, Raphaël},
  title = {Machine stability and dosimetry for ultra‐high dose rate FLASH radiotherapy human clinical protocol},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70102}
}
Kostenlos starten →