Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters

Kevin Liu, Allison Palmiero, Nitish Chopra, Brett Velasquez, Ziyi Li, Sam Beddar, Emil Schüler

Journal of Applied Clinical Medical Physics · 2023-02 · 24 (2)

DOI: 10.1002/acm2.13891 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 04.01.2023. Verfasst von Kevin Liu und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, Division of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Electrons, Radiotherapy Dosage. Die Arbeit wurde bisher 64-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate the usefulness and effectiveness of a dual beam‐current transformer (BCTs) design to monitor and record the beam dosimetry output and energy of pulsed electron FLASH (eFLASH) beams in real‐time, and to inform on the usefulness of this design for future eFLASH beam control.

Methods

Two BCTs are integrated into the head of a FLASH Mobetron system, one located after the primary scattering foil and the other downstream of the secondary scattering foil. The response of the BCTs was evaluated individually to monitor beam output as a function of dose, scattering conditions, and ability to capture physical beam parameters such as pulse width (PW), pulse repetition frequency (PRF), and dose per pulse (DPP), and in combination to determine beam energy using the ratio of the lower‐to‐upper BCT signal.

Results

A linear relationship was observed between the absorbed dose measured on Gafchromic film and the BCT signals for both the upper and lower BCT ( R 2 > 0.99). A linear relationship was also observed in the BCT signals as a function of the number of pulses delivered regardless of the PW, DPP, or PRF ( R 2 > 0.99). The lower‐to‐upper BCT ratio was found to correlate strongly with the energy of the eFLASH beam due to differential beam attenuation caused by the secondary scattering foil. The BCTs were also able to provide accurate information about the PW, PRF, energy, and DPP for each individual pulse delivered in real‐time.

Conclusion

The dual BCT system integrated within the FLASH Mobetron was shown to be a reliable monitoring system able to quantify accelerator performance and capture all essential physical beam parameters on a pulse‐by‐pulse basis, and the ratio between the two BCTs was strongly correlated with beam energy. The fast signal readout and processing enables the BCTs to provide real‐time information on beam output and energy and is proposed as a system suitable for accurate beam monitoring and control of eFLASH beams.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsRadiotherapy DosageHumansRadiometry

Stichwörter der Autor:innen

FLASHbeam current transformersdosimetrytoroids

Autor:innen und Institutionen

  • Department of Radiation Physics, Division of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Graduate School of Biomedical Sciences, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Department of Biostatistics, Division of Basic Sciences, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.

Förderung

  • NCI NIH HHS

Zitieren

APA

Liu, K., Palmiero, A., Chopra, N., Velasquez, B., Li, Z., Beddar, S., & Schüler, E. (2023). Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters. Journal of Applied Clinical Medical Physics, 24(2). https://doi.org/10.1002/acm2.13891

Vancouver

Liu K, Palmiero A, Chopra N, Velasquez B, Li Z, Beddar S, et al. Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters. Journal of Applied Clinical Medical Physics. 2023;24(2). doi:10.1002/acm2.13891

BibTeX
@article{Liu2023,
  author = {Liu, Kevin and Palmiero, Allison and Chopra, Nitish and Velasquez, Brett and Li, Ziyi and Beddar, Sam and Schüler, Emil},
  title = {Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  volume = {24},
  number = {2},
  doi = {10.1002/acm2.13891}
}
Kostenlos starten →