Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Hamid Reza Baghani und 1 weiteren Autor:innen. Beteiligte Einrichtung: Physics Department, Hakim Sabzevari University, Sabzevar, Iran. Hauptthemen laut MEDLINE: Electrons, Lead. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Objective Avoiding the underlying healthy tissue over‐exposure during breast intraoperative electron radiotherapy (IOERT) is owing to the use of some dedicated radioprotection disks during patient irradiation. The originated contaminant photons from some widely used double‐layered shielding disks including PMMA+Cu, PTFE+steel, and Al+Pb configurations during the breast IOERT have been evaluated through a Monte Carlo (MC) simulation approach.

Methods

Produced electron beam with energies of 6, 8, 10, and 12 MeV by a validated MC model of Liac12 dedicated IOERT accelerator was used for disk irradiations. Each of above‐mentioned radioprotection disks was simulated inside a water phantom, so that the upper disk surface was positioned at R 90 depth of each considered electron energy. Simulations were performed by MCNPX (version 2.6.0) MC code. Then, the energy spectra of the contaminant photons at different disk surfaces (upper, middle, and lower one) and relevant contaminant dose beneath the studied disks were determined and compared.

Results

None of studied shielding disks show significant photon contamination up to 10 MeV electron energy, so that the induced photon dose by the contaminant X‐rays was lower than those observed in the disk absence under the same conditions. In return, the induced photon dose at a close distance to the lower disk surface exceeded from calculated values in the disk absence at 12 MeV electron energy. The best performance in contaminant dose reduction at the energy range of 6–10 MeV belonged to the Al+Pb disk, while the PMMA+Cu configuration showed the best performance in this regard at 12 MeV energy.

Conclusion

Finally, it can be concluded that all studied shielding disks not only don't produce considerable photon contamination but also absorb the originated X‐rays from electron interactions with water at the electron energy range of 6–10 MeV. The only concern is related to 12 MeV energy where the induced photon dose exceeds the dose values in the disk absence. Nevertheless, the administered dose by contaminant photons to underlying healthy tissues remains beneath the tolerance dose level by these organs at the entire range of studied electron energies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsLeadHumansMonte Carlo MethodPolymethyl MethacrylatePhotonsWaterRadiometryRadiotherapy Dosage

Stichwörter der Autor:innen

Monte Carlo simulationbreast cancerelectron beamintraoperative radiotherapyphoton contaminationradioprotection disk

Substanzen

LeadPolymethyl MethacrylateWater

Autor:innen und Institutionen

  • Physics Department, Hakim Sabzevari University, Sabzevar, Iran.
  • Medical Physics and Radiological Sciences Department, Sabzevar University of Medical Sciences, Sabzevar, Iran.
  • Non-communicable Disease Research Center, Sabzevar University of Medical Sciences, Sabzevar, Iran.

Zitieren

APA

Baghani, H. R., & Robatjazi, M. (2023). Evaluating the induced photon contamination by different breast IOERT shields using Monte Carlo simulation. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14098

Vancouver

Baghani HR, Robatjazi M. Evaluating the induced photon contamination by different breast IOERT shields using Monte Carlo simulation. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14098

BibTeX
@article{Baghani2023,
  author = {Baghani, Hamid Reza and Robatjazi, Mostafa},
  title = {Evaluating the induced photon contamination by different breast IOERT shields using Monte Carlo simulation},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14098}
}
Kostenlos starten →