Dosimetric validation of intensity‐modulated bolus electron conformal therapy planning and delivery using an anthropomorphic cylindrical head phantom

Andrew S McGuffey, Garrett M Pitcher, Kevin J Erhart, Kenneth R Hogstrom

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14347 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Andrew S McGuffey und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Astronomy, Louisiana State University, Baton Rouge, Louisiana, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiotherapy, Intensity-Modulated. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, National Cancer Institute of the National Institutes of Health (NIH), 2R44CA199838-02.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This work investigated the dosimetric accuracy of the intensity‐modulated bolus electron conformal therapy (IM‐BECT) planning and delivery process using the decimal ElectronRT (eRT) treatment planning system.

Methods

An IM‐BECT treatment plan was designed using eRT for a cylindrical, anthropomorphic retromolar trigone phantom. Treatment planning involved specification of beam parameters and design of a variable thickness wax bolus and Passive Radiotherapy Intensity Modulator for Electrons (PRIME) device, which was comprised of 33 tungsten island blocks of discrete diameters from 0.158 to 0.223 cm (Intensity Reduction Factors from 0.937 to 0.875, respectively) inside a 10.1 × 6.7 cm 2 copper cutout. For comparison of calculation accuracy, a BECT plan was generated by copying the IM‐BECT plan and removing the intensity modulation. For both plans, a 16 MeV electron beam was used with 104.7 cm source‐to‐surface distance to bolus. In‐phantom TLD‐100 measurements ( N = 47) were compared with both eRT planned dose distributions, which used the pencil beam redefinition algorithm with modifications for passive electron intensity modulation (IM‐PBRA). Dose difference and distance to agreement (DTA) metrics were computed for each measurement point.

Results

Comparison of measured dose distributions with planned dose distributions yielded dose differences (calculated minus measured) characterized by a mean and standard deviation of −0.36% ± 1.64% for the IM‐BECT plan, which was similar to −0.36% ± 1.90% for the BECT plan. All dose measurements were within 5% of the planned dose distribution, with both the BECT and IM‐BECT measurement sets having 46/47 (97.8%) points within 3% or within 3 mm of the respective treatment plans.

Conclusions

It was found that the IM‐BECT treatment plan generated using eRT was sufficiently accurate for clinical use when compared to TLD measurements in a cylindrical, anthropomorphic phantom, and was similarly accurate to the BECT treatment plan in the same phantom.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiotherapy, Intensity-ModulatedElectronsHumansAlgorithmsOrgans at RiskRadiometryHead

Stichwörter der Autor:innen

electron intensity modulationelectron radiotherapyradiation measurementtreatment planning

Autor:innen und Institutionen

  • Department of Physics and Astronomy, Louisiana State University, Baton Rouge, Louisiana, USA.
  • Mary Bird Perkins Cancer Center, Baton Rouge, Louisiana, USA.
  • decimal, LLC, Sanford, Florida, USA.

Förderung

  • NCI NIH HHS
  • National Cancer Institute of the National Institutes of Health (NIH)
  • 2R44CA199838-02

Zitieren

APA

McGuffey, A. S., Pitcher, G. M., Erhart, K. J., & Hogstrom, K. R. (2024). Dosimetric validation of intensity‐modulated bolus electron conformal therapy planning and delivery using an anthropomorphic cylindrical head phantom. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14347

Vancouver

McGuffey AS, Pitcher GM, Erhart KJ, Hogstrom KR. Dosimetric validation of intensity‐modulated bolus electron conformal therapy planning and delivery using an anthropomorphic cylindrical head phantom. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14347

BibTeX
@article{McGuffey2024,
  author = {McGuffey, Andrew S and Pitcher, Garrett M and Erhart, Kevin J and Hogstrom, Kenneth R},
  title = {Dosimetric validation of intensity‐modulated bolus electron conformal therapy planning and delivery using an anthropomorphic cylindrical head phantom},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14347}
}
Kostenlos starten →