Kurz gefasst

Fachartikel in der Zeitschrift Cureus, erschienen am 08.07.2025. Verfasst von Masato Takanashi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Radiation Oncology, Tokyo Medical University Ibaraki Medical Center, Inashiki-gun, JPN. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The purpose of this study was to determine a new conversion factor (kSSDE-LP) that would allow more accurate calculation of the effective dose that is minimally influenced by patient body size, using the Dose Management Software (DMS), and to verify its accuracy in estimating patient-specific effective dose compared to existing methods. A simple regression equation was obtained using the product of the size-specific dose estimate (SSDE) and scan length on the horizontal axis and the effective dose on the vertical axis, and the slope was taken as the effective dose conversion coefficient, kSSDE-LP. Similarly, the slope obtained from another simple regression equation, using the dose-length product (DLP) on the horizontal axis and effective dose on the vertical axis, was defined as kDLP. The effective dose conversion factors and coefficients of determination (R2) were compared for males, females, and both sexes. The DLP, SSDE, scan length, and effective dose, which are dose indices necessary for determining effective dose conversion coefficients, were obtained from the DMS. The kSSDE-LP values in males, females, and both sexes were 0.012 (R2 = 0.997), 0.014 (R2 = 0.996), and 0.013 (R2 = 0.993), respectively. Using the SSDE, a dose index that takes into account information about the patient's physique, we calculated the kSSDE-LP, which reflects the current tissue weighting coefficients and mathematical voxel phantom and does not easily deviate from the regression equation even in the high-weight group.

Schlagwörter

Stichwörter der Autor:innen

compute tomographydose length productdose management softwareeffective doseradiation protectionsize-specific dose estimates

Autor:innen und Institutionen

  • Department of Radiology and Radiation Oncology, Tokyo Medical University Ibaraki Medical Center, Inashiki-gun, JPN.
  • Department of Urology, Tokyo Medical University Ibaraki Medical Center, Inashiki-gun, JPN.

Zitieren

APA

Takanashi, M., Katada, Y., Kuroda, I., Hoshina, M., Matsushita, M., Masuda, K., & Sugahara, S. (2025). Usefulness of a New Effective Dose Conversion Factor Using Size-Specific Dose Estimate (SSDE) in Abdominopelvic Region. Cureus. https://doi.org/10.7759/cureus.87499

Vancouver

Takanashi M, Katada Y, Kuroda I, Hoshina M, Matsushita M, Masuda K, et al. Usefulness of a New Effective Dose Conversion Factor Using Size-Specific Dose Estimate (SSDE) in Abdominopelvic Region. Cureus. 2025. doi:10.7759/cureus.87499

BibTeX
@article{Takanashi2025,
  author = {Takanashi, Masato and Katada, Yoshiaki and Kuroda, Isao and Hoshina, Masataka and Matsushita, Masayoshi and Masuda, Koichi and Sugahara, Shinji},
  title = {Usefulness of a New Effective Dose Conversion Factor Using Size-Specific Dose Estimate (SSDE) in Abdominopelvic Region},
  journal = {Cureus},
  year = {2025},
  doi = {10.7759/cureus.87499}
}
Kostenlos starten →