Relationship Between Effective Dose, Alternative Metrics, and SSDE: Experiences with Two CT Dose-Monitoring Systems
Diagnostics · 28.06.2025 · 15 (13) · 1654
DOI: 10.3390/diagnostics15131654 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 28.06.2025. Verfasst von Lilla Szatmáriné Egeresi und 10 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiology and Imaging Science, Department of Medical Imaging, Faculty of Medicine, University of Debrecen, Nagyerdei St. 98, H-4032 Debre…. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Culture and Innovation of Hungary from the National Research, Development and Innovation Fund.
Abstract (Originalsprache)
Background
We assessed the frequency and causes of discrepancies in CT dose indices such as dose-length product (DLP), size-specific dose estimate (SSDE), and effective dose (ED), as calculated by CT dose-monitoring systems. Our secondary aim was to demonstrate the estimation of size-specific ED (SED) from the patients’ dose records.
Methods
The retrospective study included dosimetric data of 79,383 consecutive CT exams performed on two CT scanners. The following dose values were recorded from both the locally developed dose-monitoring system (DMS) and a commercial dose-monitoring program (DWTM): DLP, SSDE, and ED. Only the DMS provided bodyweight-corrected effective dose (SEDDMS) and the SED based on previous published data.
Results
Without body-region-specific analysis, there were no tendentious differences between the DLP, ED, or SSDE values obtained from DWTM and DMS. However, the body region-based correlation revealed substantial differences between EDDMS and EDDW, primarily related to inadequate identification of the body. SSDE showed strong correlation to each anatomical category and CT device, except for the head region, where inadequate consideration of CT inclination was the reason for the biased SSDEDW value. Furthermore, by analyzing the SEDDMS, SSDE, and SED correlations, we concluded that SEDDMS is a promising figure for estimating the SED value.
Conclusions
SED provides suitable supplementary size-specific dose data to SDDE and may be a preferable choice for estimating cumulative doses in routine radiological practice.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Radiology and Imaging Science, Department of Medical Imaging, Faculty of Medicine, University of Debrecen, Nagyerdei St. 98, H-4032 Debrecen, Hungary.
- Doctoral School of Molecular Medicine, Nagyerdei St. 98, H-4032 Debrecen, Hungary.
- Doctoral School of Neuroscience, Nagyerdei St. 98, H-4032 Debrecen, Hungary.
- Iconomix Kft. Koller St. 9B 5., H-7626 Pécs, Hungary.
- Division of Nuclear Medicine and Translational Imaging, Department of Medical Imaging, Faculty of Medicine, University of Debrecen, Nagyerdei St. 98, H-4032 Debrecen, Hungary.
ORCID-Profile: Lilla Szatmáriné Egeresi, Ervin Balázs, János Kiss, Péter Bágyi, Zita Képes, Miklós Emri, László Balkay
Förderung
- Ministry of Culture and Innovation of Hungary from the National Research, Development and Innovation Fund
Zitieren
APA
Egeresi, L. S., Urbán, L., Dankó, Z., Balázs, E., Berényi, E., Marosi, M., Kiss, J., Bágyi, P., Képes, Z., Emri, M., & Balkay, L. (2025). Relationship Between Effective Dose, Alternative Metrics, and SSDE: Experiences with Two CT Dose-Monitoring Systems. Diagnostics, 15(13), 1654. https://doi.org/10.3390/diagnostics15131654
Vancouver
Egeresi LS, Urbán L, Dankó Z, Balázs E, Berényi E, Marosi M, et al. Relationship Between Effective Dose, Alternative Metrics, and SSDE: Experiences with Two CT Dose-Monitoring Systems. Diagnostics. 2025;15(13):1654. doi:10.3390/diagnostics15131654
BibTeX
@article{Egeresi2025,
author = {Egeresi, Lilla Szatmáriné and Urbán, László and Dankó, Zsolt and Balázs, Ervin and Berényi, Ervin and Marosi, Mária and Kiss, János and Bágyi, Péter and Képes, Zita and Emri, Miklós and Balkay, László},
title = {Relationship Between Effective Dose, Alternative Metrics, and SSDE: Experiences with Two CT Dose-Monitoring Systems},
journal = {Diagnostics},
year = {2025},
volume = {15},
number = {13},
pages = {1654},
doi = {10.3390/diagnostics15131654}
}