Evaluation of a new Electrochemiluminescence Immunoassay System for Measuring High-Sensitivity Cardiac Troponin T
Clinical Laboratory · 2024
DOI: 10.7754/clin.lab.2023.230640 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Laboratory, 2024. Verfasst von Jinying Wang und 7 weiteren Autor:innen. Hauptthemen laut MEDLINE: Troponin T, Immunologic Tests.
Abstract (Originalsprache)
BACKGROUND: The aim of this study was to verify the analytical performance of the UD90DT electrochemiluminescence immunoassay system (the UD90DT system) for measuring high-sensitivity cardiac troponin T (hs-cTnT). METHODS: According to the Clinical and Laboratory Standards Institute guidelines, the imprecision, linearity, reference interval, limit of blank (LoB), limit of detection (LoD), and functional sensitivity (FS) of hs-cTnT using the UD90DT system were verified. The trueness was validated using the Proficiency Testing (PT) materials. RESULTS: The within-run and between-run coefficients of variations (CVs) of two hs-cTnT levels were 7.2% and 1.5%, and 7.1% and 2.6%, respectively. The biases of the PT samples (n = 6) all fell within the allowable total error. The linearity satisfied the requirements, with a slope of 0.9963 and an R12 value of 0.9998. The hs-cTnT levels of the healthy volunteers (n = 20) ranged from 3.0 ng/L to 7.7 ng/L. All blank calibrator measurements (n = 20) fell within the LoD claim, and none of the samples (n = 25) had a LoB value ≤ 3.0 ng/L. The FS was 5.3 ng/L. Furthermore, a good correlation between the UD90DT system and the Cobas e 601 module was observed for hs-cTnT. CONCLUSIONS: The analytical performance of hs-cTnT using the UD90DT system is acceptable and satisfies clinical needs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Zitieren
APA
Wang, J., Meng, Q., Zhao, Y., Li, Z., Ling, L., Xu, Z., Li, Z., & Li, X. (2024). Evaluation of a new Electrochemiluminescence Immunoassay System for Measuring High-Sensitivity Cardiac Troponin T. Clinical Laboratory. https://doi.org/10.7754/clin.lab.2023.230640
Vancouver
Wang J, Meng Q, Zhao Y, Li Z, Ling L, Xu Z, et al. Evaluation of a new Electrochemiluminescence Immunoassay System for Measuring High-Sensitivity Cardiac Troponin T. Clinical Laboratory. 2024. doi:10.7754/clin.lab.2023.230640
BibTeX
@article{Wang2024,
author = {Wang, Jinying and Meng, Qianlong and Zhao, Yi and Li, Zhiyuan and Ling, Lu and Xu, Zhenzhen and Li, Zhen and Li, Xiaoyang},
title = {Evaluation of a new Electrochemiluminescence Immunoassay System for Measuring High-Sensitivity Cardiac Troponin T},
journal = {Clinical Laboratory},
year = {2024},
doi = {10.7754/clin.lab.2023.230640}
}
EUVIMED