Analytical Performance of the Sixth Generation High-Sensitivity Cardiac Troponin T Assay

Mengyuan Ge, Lori B Daniels iD, Spencer Seely, Ronald W McLawhon, Robert L Fitzgerald iD, Raymond T Suhandynata iD

The Journal of Applied Laboratory Medicine · 03.09.2026 · 11 (5) · 1026-1038

DOI: 10.1093/jalm/jfag055 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 09.06.2026. Verfasst von Mengyuan Ge und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pathology, University of California San Diego, La Jolla, CA, United States. Hauptthemen laut MEDLINE: Troponin T, Myocardial Infarction. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Roche Diagnostics, University of California and UC San Diego Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

High-sensitivity cardiac troponin assays are integral to diagnosing myocardial infarction. The current Roche Elecsys® Troponin T high-sensitivity Gen 5 assay (Gen 5) received FDA clearance in 2017. A sixth-generation assay was recently developed with improved analytical characteristics, including sensitivity and resistance to interferences. Reference ranges for this assay were reported, but analytical performance in the clinical laboratory has yet to be evaluated.

Methods

The Elecsys® Troponin T high-sensitivity Gen 6 (Gen 6) assay was evaluated on the Roche cobas e 801 analyzers. Measurement limits of the assay were determined in lithium-heparin plasma specimens per CLSI EP17-A2. Repeatability and intermediate precision of QC materials and plasma specimens were determined according to CLSI EP05-A2 and EP15-A2, respectively.

Results

Serial dilutions of cardiac troponin T revealed that Gen 6 exhibits a positive bias relative to Gen 5 at concentrations >10 ng/L. Below this point, the bias reverses, indicating concentration-dependent crossover. The limit of blank and limit of detection were 0.4 ng/L and 1.2 ng/L, respectively. The limit of quantitation was 2.2 ng/L with a 10% CV. No cross-reactivity was observed on Gen 6 for skeletal muscle troponin T, cardiac troponin I, skeletal muscle troponin I, or troponin C at concentrations up to 10 000 ng/L. Interference from hemolysis was not observed up to 5000 mg/dL of hemoglobin.

Conclusions

The Gen 6 assay demonstrates resistance to interference and greater analytical sensitivity, with a lower limit of detection and limit of quantitation relative to Gen 5, enabling broader coverage of healthy individuals.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Troponin TMyocardial InfarctionHumansReproducibility of ResultsLimit of DetectionReference ValuesSensitivity and SpecificityBiomarkersImmunoassay

Substanzen

Troponin TBiomarkers

Autor:innen und Institutionen

  • Department of Pathology, University of California San Diego, La Jolla, CA, United States.
  • Division of Cardiovascular Medicine, University of California San Diego, La Jolla, CA, United States.
  • Skaggs School of Pharmacy and Pharmaceutical Sciences, University of California San Diego, La Jolla, CA, United States.

ORCID-Profile: Lori B Daniels, Robert L Fitzgerald, Raymond T Suhandynata

Förderung

  • Roche Diagnostics
  • University of California and UC San Diego Health

Zitieren

APA

Ge, M., Daniels, L. B., Seely, S., McLawhon, R. W., Fitzgerald, R. L., & Suhandynata, R. T. (2026). Analytical Performance of the Sixth Generation High-Sensitivity Cardiac Troponin T Assay. The Journal of Applied Laboratory Medicine, 11(5), 1026-1038. https://doi.org/10.1093/jalm/jfag055

Vancouver

Ge M, Daniels LB, Seely S, McLawhon RW, Fitzgerald RL, Suhandynata RT. Analytical Performance of the Sixth Generation High-Sensitivity Cardiac Troponin T Assay. The Journal of Applied Laboratory Medicine. 2026;11(5):1026-1038. doi:10.1093/jalm/jfag055

BibTeX
@article{Ge2026,
  author = {Ge, Mengyuan and Daniels, Lori B and Seely, Spencer and McLawhon, Ronald W and Fitzgerald, Robert L and Suhandynata, Raymond T},
  title = {Analytical Performance of the Sixth Generation High-Sensitivity Cardiac Troponin T Assay},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2026},
  volume = {11},
  number = {5},
  pages = {1026--1038},
  doi = {10.1093/jalm/jfag055}
}
Kostenlos starten →