Stress distribution analysis of miniplates used in mandibular fracture fixation: An in vitro study
Bioinformation · 31.07.2026 · 22 (7) · 4153-4158
DOI: 10.6026/973206300224153 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioinformation, online erschienen am 31.07.2026. Verfasst von Jiwanasha Manish Agrawal.
Abstract (Originalsprache)
Achieving optimal mechanical stability while minimizing stress concentration remains a major challenge in the fixation of mandibular angle fractures. Therefore, it is of interest to evaluate the stress distribution and fracture segment displacement of four miniplate fixation designs using 48 standardized polyurethane mandibular angle fracture models subjected to vertical loading. Three dimensional strut miniplates showed the lowest peak strain (812.7 ą 93.8 με), least fracture-gap displacement (0.32 ą 0.05 mm) and highest load resistance at 1-mm displacement (274.6 ą 31.9 N), significantly outperforming the other fixation systems (p
Zitieren
APA
Agrawal, J. M. (2026). Stress distribution analysis of miniplates used in mandibular fracture fixation: An in vitro study. Bioinformation, 22(7), 4153-4158. https://doi.org/10.6026/973206300224153
Vancouver
Agrawal JM. Stress distribution analysis of miniplates used in mandibular fracture fixation: An in vitro study. Bioinformation. 2026;22(7):4153-4158. doi:10.6026/973206300224153