EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Screw Stress Distribution in a Clavicle Fracture with Plate Fixation: A Finite Element Analysis

Angelo Alito iD, Domenico Fenga iD, Giada Tropeano, Demetrio Milardi, Danilo Leonetti, Alba Migliorato, Adriana Tisano, Danilo D'Andrea iD, Vincenzo Filardi

Bioengineering · 07.12.2023 · 10 (12) · 1402

DOI: 10.3390/bioengineering10121402 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 07.12.2023. Verfasst von Angelo Alito und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical, Dental Sciences and Morphological and Functional Images, University of Messina, 98125 Messina, Italy. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 78 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Clavicle midshaft fractures are mostly treated surgically by open internal reduction with a superior or anteroinferior plate and screws or by intramedullary nailing. Screw positioning plays a critical role in determining the stress distribution. There is a lack of data on the screw position and the appropriate number of cortices required for plate fixation. The aim of this study is to evaluate the mechanical behavior of an anterior plate implanted in a fractured bone subjected to 120° of lateral elevation compared to a healthy clavicle using numerical simulations. Contact forces and moments used were obtained from literature data and applied to the healthy and fractured finite element models. Stresses of about 9 MPa were found on the healthy clavicle, while values of about 15 MPa were calculated on the plate of the fractured one; these stress peaks were reached at about 30° and 70° of elevation when the stress shielding on the clavicle sums all the three components of the solicitation: compression, flexion, and torsion. The stress distribution in a clavicle fracture stabilized with plates and screws is influenced by several factors, including the plate’s position and design, the type of screw, and the biomechanical forces applied during movements.

Schlagwörter

Stichwörter der Autor:innen

clavicle platefinite element analysisfinite element modelingfractured claviclescrew stress distributionstress analysis

Autor:innen und Institutionen

  • Department of Biomedical, Dental Sciences and Morphological and Functional Images, University of Messina, 98125 Messina, Italy.
  • Department of Orthopaedics and Traumatology, University Hospital A.O.U. "G. Martino", 98125 Messina, Italy.
  • Department of Clinical and Experimental Medicine, University of Messina, 98125 Messina, Italy.
  • Department of Engineering, University of Messina, 98158 Messina, Italy.
  • D.A. Scientific Research and Internationalization, University of Messina, 98121 Messina, Italy.

ORCID-Profile: Angelo Alito, Domenico Fenga, Danilo D'Andrea

Zitieren

APA

Alito, A., Fenga, D., Tropeano, G., Milardi, D., Leonetti, D., Migliorato, A., Tisano, A., D'Andrea, D., & Filardi, V. (2023). Screw Stress Distribution in a Clavicle Fracture with Plate Fixation: A Finite Element Analysis. Bioengineering, 10(12), 1402. https://doi.org/10.3390/bioengineering10121402

Vancouver

Alito A, Fenga D, Tropeano G, Milardi D, Leonetti D, Migliorato A, et al. Screw Stress Distribution in a Clavicle Fracture with Plate Fixation: A Finite Element Analysis. Bioengineering. 2023;10(12):1402. doi:10.3390/bioengineering10121402

BibTeX
@article{Alito2023,
  author = {Alito, Angelo and Fenga, Domenico and Tropeano, Giada and Milardi, Demetrio and Leonetti, Danilo and Migliorato, Alba and Tisano, Adriana and D'Andrea, Danilo and Filardi, Vincenzo},
  title = {Screw Stress Distribution in a Clavicle Fracture with Plate Fixation: A Finite Element Analysis},
  journal = {Bioengineering},
  year = {2023},
  volume = {10},
  number = {12},
  pages = {1402},
  doi = {10.3390/bioengineering10121402}
}
Kostenlos starten →