Evaluation of the Distribution of Spatial Scattered Radiation Dose in Bedside Dynamic Chest Radiography

Kazuki Otomo, Tomoyoshi Kawabata, Syu Onodera

Japanese Journal of Radiological Technology · 2025

DOI: 10.6009/jjrt.25-1471 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Japanese Journal of Radiological Technology, 2025. Verfasst von Kazuki Otomo und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology Division of Medical Technology, Tohoku University Hospital. Hauptthemen laut MEDLINE: Radiography, Thoracic, Phantoms, Imaging, Radiation Dosage. Die Arbeit nennt 15 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The purpose of this study was to determine the distribution of spatial scattered radiation during mobile dynamic chest radiography and to develop protective measures. METHODS: We measured the spatial scattered radiation, assuming both mobile chest radiography and mobile dynamic chest radiography. We placed the chest phantom on a stretcher. Gatch-up angles were varied from 75°(seated position), 35°(semi-seated position), and 0°(supine position). We measured spatial scattered radiation using the ionization chamber survey meter. The height of the measurement points was 100 cm and 150 cm, assuming the position of the abdomen and lens. We also measured spatial scattered radiation using the radiation shielding screen. RESULTS: Spatial scattered radiation dose in mobile dynamic chest radiography was approximately five to six times higher than in mobile chest radiography. In the seated position, the spatial scattered radiation dose was about 1.5 times higher on the caudal side of the phantom compared to the cranial side. Behind the radiation shielding screen, the spatial scattered radiation dose decreased significantly. CONCLUSION: If the examination is performed in a seated position, the radiological technologist can reduce exposure by moving to the patient's head side. Radiation shielding screens are very effective in reducing radiation exposure in mobile dynamic chest radiography.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiography, ThoracicPhantoms, ImagingRadiation DosageHumansScattering, RadiationRadiation Protection

Stichwörter der Autor:innen

digital radiographydosimetrymobile X-ray unitradiation protectionscattered radiation

Autor:innen und Institutionen

  • Department of Radiology Division of Medical Technology, Tohoku University Hospital.

Zitieren

APA

Otomo, K., Kawabata, T., & Onodera, S. (2025). Evaluation of the Distribution of Spatial Scattered Radiation Dose in Bedside Dynamic Chest Radiography. Japanese Journal of Radiological Technology. https://doi.org/10.6009/jjrt.25-1471

Vancouver

Otomo K, Kawabata T, Onodera S. Evaluation of the Distribution of Spatial Scattered Radiation Dose in Bedside Dynamic Chest Radiography. Japanese Journal of Radiological Technology. 2025. doi:10.6009/jjrt.25-1471

BibTeX
@article{Otomo2025,
  author = {Otomo, Kazuki and Kawabata, Tomoyoshi and Onodera, Syu},
  title = {Evaluation of the Distribution of Spatial Scattered Radiation Dose in Bedside Dynamic Chest Radiography},
  journal = {Japanese Journal of Radiological Technology},
  year = {2025},
  doi = {10.6009/jjrt.25-1471}
}
Kostenlos starten →