Effect of radioprotective curtain length on the scattered dose rate distribution and endoscopist eye lens dose with an over-couch fluoroscopy system

Kosuke Matsubara, Asuka Nakajima, Ayaka Hirosawa, Ryo Yoshikawa, Nao Ichikawa, Kotaro Fukushima, Atsushi Fukuda

Physical and Engineering Sciences in Medicine · 2024

DOI: 10.1007/s13246-024-01398-w ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2024. Verfasst von Kosuke Matsubara und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, …. Hauptthemen laut MEDLINE: Lens, Crystalline, Radiation Protection, Radiation Dosage, Monte Carlo Method. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Society for the Promotion of Science, Kanazawa University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Sufficient dose reduction may not be achieved if radioprotective curtains are folded. This study aimed to evaluate the scattered dose rate distribution and physician eye lens dose at different curtain lengths. Using an over-couch fluoroscopy system, d H *(10)/d t was measured using a survey meter 150 cm from the floor at 29 positions in the examination room when the curtain lengths were 0% (no curtain), 50%, 75%, and 100%. The absorbed dose rates in the air at the positions of endoscopist and assistant were calculated using a Monte Carlo simulation by varying the curtain length from 0 to 100%. The air kerma was measured by 10 min fluoroscopy using optically stimulated luminescence dosimeters at the eye surfaces of the endoscopist phantom and the outside and inside of the radioprotective goggles. At curtain lengths of 50%, 75%, and 100%, the ratios of d H *(10)/d t relative to 0% ranged from 80.8 to 104.1%, 10.5 to 61.0%, and 11.8 to 24.8%, respectively. In the simulation, the absorbed dose rates at the endoscopist’s and assistant’s positions changed rapidly between 55 and 75% and 65% and 80% of the curtain length, respectively. At the 0%, 50%, 75%, and 100% curtain lengths, the air kerma at the left eye surface of the endoscopist phantom was 237 ± 29, 271 ± 30, 37.7 ± 7.5, and 33.5 ± 6.1 μGy, respectively. Therefore, a curtain length of 75% or greater is required to achieve a sufficient eye lens dose reduction effect at the position of the endoscopist.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lens, CrystallineRadiation ProtectionRadiation DosageMonte Carlo MethodScattering, RadiationFluoroscopyHumansPhantoms, ImagingDose-Response Relationship, Radiation

Stichwörter der Autor:innen

Eye lensOccupational exposureRadiation measurementRadiation protectionRadioprotective curtain

Autor:innen und Institutionen

  • Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan. matsuk@mhs.mp.…
  • Department of Radiological Technology, School of Health Sciences, College of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan.
  • Division of Radiology, Shinshu University Hospital, 3-1-1 Asahi, Matsumoto, Nagano, 390-8621, Japan.
  • Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan.
  • Department of Medical Technology, Toyama Prefectural Central Hospital, 2-2-78 Nishinagae, Toyoma, Toyama, 930-8550, Japan.
  • Department of Radiology, Kanazawa University Hospital, 13-1 Takaramachi, Kanazawa, Ishikawa, 920-8641, Japan.

Förderung

  • Japan Society for the Promotion of Science
  • Kanazawa University

Zitieren

APA

Matsubara, K., Nakajima, A., Hirosawa, A., Yoshikawa, R., Ichikawa, N., Fukushima, K., & Fukuda, A. (2024). Effect of radioprotective curtain length on the scattered dose rate distribution and endoscopist eye lens dose with an over-couch fluoroscopy system. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-024-01398-w

Vancouver

Matsubara K, Nakajima A, Hirosawa A, Yoshikawa R, Ichikawa N, Fukushima K, et al. Effect of radioprotective curtain length on the scattered dose rate distribution and endoscopist eye lens dose with an over-couch fluoroscopy system. Physical and Engineering Sciences in Medicine. 2024. doi:10.1007/s13246-024-01398-w

BibTeX
@article{Matsubara2024,
  author = {Matsubara, Kosuke and Nakajima, Asuka and Hirosawa, Ayaka and Yoshikawa, Ryo and Ichikawa, Nao and Fukushima, Kotaro and Fukuda, Atsushi},
  title = {Effect of radioprotective curtain length on the scattered dose rate distribution and endoscopist eye lens dose with an over-couch fluoroscopy system},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2024},
  doi = {10.1007/s13246-024-01398-w}
}
Kostenlos starten →