Evaluation on the Availability of the Electrometer Inspection by Users Using a Newly Developed Current Source

Hayato Tsuno, Fumiyasu Matsubayashi, Koji Sasaki, Takashi Sakai, Keiji Matsumoto, Kiyosh Takeuchi

Japanese Journal of Radiological Technology · 2023 · 79 (2) · 166-176

DOI: 10.6009/jjrt.2023-1216 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Japanese Journal of Radiological Technology, 2023. Verfasst von Hayato Tsuno und 5 weiteren Autor:innen. Beteiligte Einrichtung: Gunma Prefectural College of Health Sciences Graduate School of Radiological Technology. Hauptthemen laut MEDLINE: Particle Accelerators, Radiometry. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 10 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: If we try to perform the inspection for an electrometer, two types of electronic signals energized to the electrometer can be used. One is the signal that occurs in the ionization chamber irradiated by radiation. The other is the signal that is derived from a current source. The former signal is changed by radiation output, so we need to use two or more sets of the chambers and the electrometers in the inspection. In addition, the high-performance current source is relatively expensive. Therefore, it is difficult for users to inspect the electrometer simply. To deal with these, we have developed a new current source that allows users to perform highly accurate electrometer inspections at their own facilities. AIM: The purpose of this study was to verify that users can perform electrometer inspections with high accuracy in their own facilities by using a new current source. MATERIALS AND METHODS: A newly current source equips a dry cell battery as a charge generator. Current, polarity, and energized time can be changed by the source setting. We conducted an inspection by energizing the electrometer using the developed current source. RESULTS: The coefficient of variation of the charge amount in the measurement using the new current source was within 0.05%. The electrometer calibration coefficients measured by sensitivity comparison using this current source could be obtained based on electrometers calibrated at a certified facility of the Japan Calibration Service System. CONCLUSION: We have shown that the new current source can be used as a relative current value by using a calibrated electrometer as a reference, indicating that the user can check the electrometer.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Particle AcceleratorsRadiometryCalibrationJapan

Stichwörter der Autor:innen

calibration separatelycurrent sourcedosimeterelectrometerradiotherapy

Autor:innen und Institutionen

  • Gunma Prefectural College of Health Sciences Graduate School of Radiological Technology.
  • Japanese Foundation for Cancer Research.
  • Kawaguchi Electric Works.

Zitieren

APA

Tsuno, H., Matsubayashi, F., Sasaki, K., Sakai, T., Matsumoto, K., & Takeuchi, K. (2023). Evaluation on the Availability of the Electrometer Inspection by Users Using a Newly Developed Current Source. Japanese Journal of Radiological Technology, 79(2), 166-176. https://doi.org/10.6009/jjrt.2023-1216

Vancouver

Tsuno H, Matsubayashi F, Sasaki K, Sakai T, Matsumoto K, Takeuchi K. Evaluation on the Availability of the Electrometer Inspection by Users Using a Newly Developed Current Source. Japanese Journal of Radiological Technology. 2023;79(2):166-176. doi:10.6009/jjrt.2023-1216

BibTeX
@article{Tsuno2023,
  author = {Tsuno, Hayato and Matsubayashi, Fumiyasu and Sasaki, Koji and Sakai, Takashi and Matsumoto, Keiji and Takeuchi, Kiyosh},
  title = {Evaluation on the Availability of the Electrometer Inspection by Users Using a Newly Developed Current Source},
  journal = {Japanese Journal of Radiological Technology},
  year = {2023},
  volume = {79},
  number = {2},
  pages = {166--176},
  doi = {10.6009/jjrt.2023-1216}
}
Kostenlos starten →