EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Dosimetric impact of calibration coefficients determined using linear accelerator photon and electron beams for ionization chamber in an on-site dosimetry audit

Kensuke Tani, Akihisa Wakita, Naoki Tohyama, Yukio Fujita

Journal of Radiation Research · 2024

DOI: 10.1093/jrr/rrae054 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Radiation Research, 2024. Verfasst von Kensuke Tani und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Medical Physics, EuroMediTech Co., Ltd, 2-20-4 Higashi-Gotanda, Shinagawa, Tokyo 141-0022, Japan. Hauptthemen laut MEDLINE: Photons, Particle Accelerators, Electrons, Radiometry. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aimed to clarify the dosimetric impact of calibration beam quality for calibration coefficients of the absorbed dose to water for an ionization chamber in an on-site dosimetry audit. Institution-measured doses of 200 photon and 184 electron beams were compared with the measured dose using one year data before and after the calibration of the ionization chamber used. For photon and electron reference dosimetry, the agreements of the institution-measured dose against two measured doses in this audit were evaluated using the calibration coefficients determined using 60Co (${N}_{D,\mathrm{w},{}^{60}\mathrm{Co}}$) and linear accelerator (linac) (${N}_{D,\mathrm{w},Q}$) beams. For electron reference dosimetry, the agreement of two institution-measured doses against the measured dose was evaluated using${N}_{D,\mathrm{w},Q}$. Institution-measured doses were evaluated using direct- and cross-calibration coefficients. For photon reference dosimetry, the mean differences and standard deviation (SD) of institution-measured dose against the measured dose using ${N}_{D,\mathrm{w},{}^{60}\mathrm{Co}}$ and ${N}_{D,\mathrm{w},Q}$ were −0.1% ± 0.4% and −0.3% ± 0.4%, respectively. For electron reference dosimetry, the mean differences and SD of institution-measured dose using the direct-calibration coefficient against the measured dose using ${N}_{D,\mathrm{w},{}^{60}\mathrm{Co}}$ and ${N}_{D,\mathrm{w},Q}$ were 1.3% ± 0.8% and 0.8% ± 0.8%, respectively. Further, the mean differences and SD of institution-measured dose using the cross-calibration coefficient against the measured dose using ${N}_{D,\mathrm{w},Q}$ were −0.1% ± 0.6%. For photon beams, the dosimetric impact of introducing calibration coefficients determined using linac beams was small. For electron beams, it was larger, and the measured dose using ${N}_{D,\mathrm{w},Q}$ was most consistent with the institution-measured dose, which was evaluated using a cross-calibration coefficient.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhotonsParticle AcceleratorsElectronsRadiometryCalibrationReproducibility of ResultsRadiotherapy DosageHumans

Stichwörter der Autor:innen

auditcalibration coefficientionization chamberplastic phantom

Autor:innen und Institutionen

  • Division of Medical Physics, EuroMediTech Co., Ltd, 2-20-4 Higashi-Gotanda, Shinagawa, Tokyo 141-0022, Japan.
  • Department of Health Sciences, Komazawa University, 1-23-1 Komazawa, Setagaya, Tokyo 154-8525, Japan.

Zitieren

APA

Tani, K., Wakita, A., Tohyama, N., & Fujita, Y. (2024). Dosimetric impact of calibration coefficients determined using linear accelerator photon and electron beams for ionization chamber in an on-site dosimetry audit. Journal of Radiation Research. https://doi.org/10.1093/jrr/rrae054

Vancouver

Tani K, Wakita A, Tohyama N, Fujita Y. Dosimetric impact of calibration coefficients determined using linear accelerator photon and electron beams for ionization chamber in an on-site dosimetry audit. Journal of Radiation Research. 2024. doi:10.1093/jrr/rrae054

BibTeX
@article{Tani2024,
  author = {Tani, Kensuke and Wakita, Akihisa and Tohyama, Naoki and Fujita, Yukio},
  title = {Dosimetric impact of calibration coefficients determined using linear accelerator photon and electron beams for ionization chamber in an on-site dosimetry audit},
  journal = {Journal of Radiation Research},
  year = {2024},
  doi = {10.1093/jrr/rrae054}
}
Kostenlos starten →