Probe Heating Method for the Analysis of Solid Samples Using a Portable Mass Spectrometer

Shun Kumano, Masuyuki Sugiyama, Masuyoshi Yamada, Kazushige Nishimura, Hideki Hasegawa, Hidetoshi Morokuma, Hiroyuki Inoue, Yuichiro Hashimoto

Mass Spectrometry · 2015

DOI: 10.5702/massspectrometry.a0038 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Mass Spectrometry, 2015. Verfasst von Shun Kumano und 7 weiteren Autor:innen. Beteiligte Einrichtung: Hitachi, Ltd., Central Research Laboratory. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We previously reported on the development of a portable mass spectrometer for the onsite screening of illicit drugs, but our previous sampling system could only be used for liquid samples. In this study, we report on an attempt to develop a probe heating method that also permits solid samples to be analyzed using a portable mass spectrometer. An aluminum rod is used as the sampling probe. The powdered sample is affixed to the sampling probe or a droplet of sample solution is placed on the tip of the probe and dried. The probe is then placed on a heater to vaporize the sample. The vapor is then introduced into the portable mass spectrometer and analyzed. With the heater temperature set to 130°C, the developed system detected 1 ng of methamphetamine, 1 ng of amphetamine, 3 ng of 3,4-methylenedioxymethamphetamine, 1 ng of 3,4-methylenedioxyamphetamine, and 0.3 ng of cocaine. Even from mixtures consisting of clove powder and methamphetamine powder, methamphetamine ions were detected by tandem mass spectrometry. The developed probe heating method provides a simple method for the analysis of solid samples. A portable mass spectrometer incorporating this method would thus be useful for the onsite screening of illicit drugs.

Schlagwörter

Stichwörter der Autor:innen

illicit drugonsite screeningportable mass spectrometersolid sample analysis

Autor:innen und Institutionen

  • Hitachi, Ltd., Central Research Laboratory.
  • Hitachi High-Technologies Corp.
  • National Research Institute of Police Science.

Zitieren

APA

Kumano, S., Sugiyama, M., Yamada, M., Nishimura, K., Hasegawa, H., Morokuma, H., Inoue, H., & Hashimoto, Y. (2015). Probe Heating Method for the Analysis of Solid Samples Using a Portable Mass Spectrometer. Mass Spectrometry. https://doi.org/10.5702/massspectrometry.a0038

Vancouver

Kumano S, Sugiyama M, Yamada M, Nishimura K, Hasegawa H, Morokuma H, et al. Probe Heating Method for the Analysis of Solid Samples Using a Portable Mass Spectrometer. Mass Spectrometry. 2015. doi:10.5702/massspectrometry.a0038

BibTeX
@article{Kumano2015,
  author = {Kumano, Shun and Sugiyama, Masuyuki and Yamada, Masuyoshi and Nishimura, Kazushige and Hasegawa, Hideki and Morokuma, Hidetoshi and Inoue, Hiroyuki and Hashimoto, Yuichiro},
  title = {Probe Heating Method for the Analysis of Solid Samples Using a Portable Mass Spectrometer},
  journal = {Mass Spectrometry},
  year = {2015},
  doi = {10.5702/massspectrometry.a0038}
}
Kostenlos starten →