Portable Sensing Probe for Real-Time Quantification of Ammonia in Blood Samples
ACS Applied Materials & Interfaces · 11.09.2024 · 16 (36) · 47242-47256
DOI: 10.1021/acsami.4c09573 ↗Kurz gefasst
Fachartikel in der Zeitschrift ACS Applied Materials & Interfaces, online erschienen am 27.08.2024. Verfasst von Rajesh Madhuvilakku und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physical Therapy, College of Healthcare Medical Science & Engineering, Inje University, Gimhae 50834, Republic of Korea. Hauptthemen laut MEDLINE: Ammonia, Colorimetry, Gold, Metal Nanoparticles. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 55 Literaturangaben. Gefördert von: National Research Foundation of Korea.
Abstract (Originalsprache)
The detection of ammonia levels in blood is critical for diagnosing and monitoring various medical conditions, including liver dysfunction and metabolic disorders. However, traditional diagnostic methods are slow and cumbersome, often involving multiple contact-based steps such as ammonia separation in alkali conditions followed by distillation or microdiffusion, leading to delays in diagnosis and treatment. Herein, we developed a colorimetric assay capable of rapid detection of ammonia in whole blood or plasma samples, utilizing 2,2,6,6-tetramethylpiperidine 1-oxyl (TEMPO)-oxidized cellulose nanocrystals (TCNC) coupled with gold nanoparticles (AuNPs). The basis of our assay relies on either (i) the interaction between the carboxylate group (-COO) of TEMPO and ammonium ions or (ii) the manipulation of AuNPs surface plasmon resonance (SPR) through the formation of Au(NH3)43+, which displaces a redox mediator, resazurin, resulting in observable multicolor displays at various concentrations of ammonia. The colorimetric assay exhibits a wide linear detection range for dissolved NH4+ (0.1-37 μM) with a low limit of detection (LOD) of 0.1 μM. Additionally, it effectively measures NH3(g) concentrations in the range of 0.5-144 μM. The fabricated electrochemical nose (E-nose) device demonstrates excellent analytical performance for plasma ammonia sensing (0.05-256 μM). Experimental results demonstrate a linear detection range suitable for clinical applications, with excellent correlation to standard laboratory methods, offering a practical solution for point-of-care (PoC) testing. We anticipate that this approach can be applied broadly to improve patient monitoring and treatment by providing immediate and accurate ammonia measurements in a clinical setting.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Physical Therapy, College of Healthcare Medical Science & Engineering, Inje University, Gimhae 50834, Republic of Korea.
- Biohealth Products Research Center (BPRC), Inje University, Gimhae 50834, Republic of Korea.
- Research Center for Aged-Life Redesign (RCAR), Inje University, Gimhae 50834, Republic of Korea.
- Department of Rehabilitation Science, Graduate School of Inje University, Gimhae 50834, Republic of Korea.
ORCID-Profile: Rajesh Madhuvilakku, Yonggeun Hong
Förderung
- National Research Foundation of Korea
Zitieren
APA
Madhuvilakku, R., & Hong, Y. (2024). Portable Sensing Probe for Real-Time Quantification of Ammonia in Blood Samples. ACS Applied Materials & Interfaces, 16(36), 47242-47256. https://doi.org/10.1021/acsami.4c09573
Vancouver
Madhuvilakku R, Hong Y. Portable Sensing Probe for Real-Time Quantification of Ammonia in Blood Samples. ACS Applied Materials & Interfaces. 2024;16(36):47242-47256. doi:10.1021/acsami.4c09573
BibTeX
@article{Madhuvilakku2024,
author = {Madhuvilakku, Rajesh and Hong, Yonggeun},
title = {Portable Sensing Probe for Real-Time Quantification of Ammonia in Blood Samples},
journal = {ACS Applied Materials & Interfaces},
year = {2024},
volume = {16},
number = {36},
pages = {47242--47256},
doi = {10.1021/acsami.4c09573}
}
EUVIMED