Optimizing dosimetric accuracy in radioembolization: comparative effectiveness of SPECT/CT, planar imaging, and PET/CT for lung shunt fraction assessment
Nuclear Medicine Review · 2024-03-26 · 27
DOI: 10.5603/nmr.102284 ↗At a glance
Vergleichsstudie in Nuclear Medicine Review, published online on 2024-03-26. Written by Subhash Chand Kheruka and 9 co-authors. Institution: Department of Radiology & Nuclear Medicine, Sultan Qaboos Comprehensive Cancer Care, and Research Center, Muscat, Oman. skheruka@gmail.com. Major MEDLINE topics: Embolization, Therapeutic, Positron Emission Tomography Computed Tomography, Radiometry. Cited 2 times so far (Crossref).
Abstract (original language)
BACKGROUND: In radioembolization therapy for hepatic malignancies, the accurate estimation of lung shunt fraction (LSF) is crucial to minimize the risk of radiation-induced pneumonitis and fibrosis due to hepatopulmonary shunting of yttrium-90 (90Y)-microspheres. This study aimed to compare the accuracy and precision of LSF estimation using technetium-99m macroaggregated albumin single photon emission computed tomography ([99mTc]Tc-MAA SPECT) LSF, [99mTc]Tc-MAA planar LSF, and 90Y PET LSF in patients undergoing 90Y-radioembolization. MATERIAL AND METHODS: A retrospective study was conducted involving 15 patients diagnosed with hepatocellular carcinoma (HCC) or liver metastases and planned to undergo transarterial radioembolization with 90Y SirSpheres after multidisplinary team discussion. LSF values were calculated using [99mTc]Tc-MAA SPECT LSF, [99mTc]Tc-MAA planar LSF, and 90Y PET LSF. The accuracy of these methods was assessed through paired t-tests and correlation analysis. RESULTS: The paired t-test revealed a statistically significant difference between SPECT LSF and planar LSF (t-statistic = -4.81, p-value = 0.0003), indicating that planar imaging tends to overestimate LSF values. However, no significant difference was observed between [99mTc]Tc-MAA SPECT LSF and 90Y PET LSF (t-statistic = -0.98, p-value = 0.343), suggesting a high degree of agreement between these two methods. Correlation analysis showed a very strong positive correlation between [99mTc]Tc-MAA SPECT LSF and 90Y PET LSF (r = 0.999), while strong correlations were also found between SPECT LSF and planar LSF, and between planar LSF and 90Y PET LSF (r = 0.841). CONCLUSIONS: The findings suggest that 90Y PET LSF aligns closely with [99mTc]Tc-MAA SPECT LSF, making them both reliable for LSF estimation in radioembolization therapy. In contrast, planar imaging tends to overestimate LSF, potentially leading to inaccurate dosimetric planning. Incorporating [99mTc]Tc-MAA SPECT/CT and 90Y PET/CT into routine clinical practice could enhance the accuracy of LSF estimation, thereby improving patient outcomes. Further research with larger cohorts is recommended to validate these findings.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Radiology & Nuclear Medicine, Sultan Qaboos Comprehensive Cancer Care, and Research Center, Muscat, Oman. skheruka@gmail.com.
- Department of Radiology & Nuclear Medicine, Sultan Qaboos Comprehensive Cancer Care, and Research Center, Muscat, Oman.
ORCID profiles: Subhash Chand Kheruka
Cite
APA
Kheruka, S. C., ALSukaiti, R., Jain, A., Al Riyami, K., Al-Balushi, A., Al-Maymani, N., Al-Makhmari, N., Al-Saidi, H., Al-Rashdi, S., & Jayakrishnan, V. (2024). Optimizing dosimetric accuracy in radioembolization: comparative effectiveness of SPECT/CT, planar imaging, and PET/CT for lung shunt fraction assessment. Nuclear Medicine Review, 27(0). https://doi.org/10.5603/nmr.102284
Vancouver
Kheruka SC, ALSukaiti R, Jain A, Al Riyami K, Al-Balushi A, Al-Maymani N, et al. Optimizing dosimetric accuracy in radioembolization: comparative effectiveness of SPECT/CT, planar imaging, and PET/CT for lung shunt fraction assessment. Nuclear Medicine Review. 2024;27(0). doi:10.5603/nmr.102284
BibTeX
@article{Kheruka2024,
author = {Kheruka, Subhash Chand and ALSukaiti, Rashid and Jain, Anjali and Al Riyami, Khulood and Al-Balushi, Anas and Al-Maymani, Naema and Al-Makhmari, Noura and Al-Saidi, Huda and Al-Rashdi, Sana and Jayakrishnan, Vipin},
title = {Optimizing dosimetric accuracy in radioembolization: comparative effectiveness of SPECT/CT, planar imaging, and PET/CT for lung shunt fraction assessment},
journal = {Nuclear Medicine Review},
year = {2024},
volume = {27},
number = {0},
doi = {10.5603/nmr.102284}
}
EUVIMED