EUVIMEDEUVIMED entdecken
Crossref

Determination of the source dwell position of an afterloading device with a detector array

Ferdinand Breithuth, Johann Schiefer, Markus Arn, Samuel Peters, Wolf Seelentag

Journal of Radiation Oncology Informatics · 2017

DOI: 10.5166/jroi-3-2-18 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Radiation Oncology Informatics, 2017. Verfasst von Ferdinand Breithuth und 4 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

It was the aim to develop and test a measurement technique for the source position of an afterloading system using an electronic two-dimensional detector array (2D array). A "GammaMed plus IX" high dose rate afterloading device (Varian Medical Systems, Palo Alto, USA), and a Seven29 2D detector array (PTW Freiburg GmbH, Freiburg, Germany) have been used. A hollow needle has been connected to the afterloading device. Its outer diameter is 3.0 mm and its length about 220mm. A 14 mm thick Perspex slab was fixed in a reproducible position on the 2D array. Above the 14th detector row, a groove was cut in the slab which accommodates the hollow needle in a fixed position relative to the 2D array. In order to define the position of the source, the signals of the detectors in the 14th detector row have been evaluated. A theoretical curve depending on the source position and strength, has been fitted to the acquired detector signals. It has been shown that the reproducibility of the measurement technique - including the reproducibility of the source placement - was within 0.15mm. This method can not only replace film measurements, it is also more exact and less time consuming.

Zitieren

APA

Breithuth, F., Schiefer, J., Arn, M., Peters, S., & Seelentag, W. (2017). Determination of the source dwell position of an afterloading device with a detector array. Journal of Radiation Oncology Informatics. https://doi.org/10.5166/jroi-3-2-18

Vancouver

Breithuth F, Schiefer J, Arn M, Peters S, Seelentag W. Determination of the source dwell position of an afterloading device with a detector array. Journal of Radiation Oncology Informatics. 2017. doi:10.5166/jroi-3-2-18

BibTeX
@article{Breithuth2017,
  author = {Breithuth, Ferdinand and Schiefer, Johann and Arn, Markus and Peters, Samuel and Seelentag, Wolf},
  title = {Determination of the source dwell position of an afterloading device with a detector array},
  journal = {Journal of Radiation Oncology Informatics},
  year = {2017},
  doi = {10.5166/jroi-3-2-18}
}
Kostenlos starten →