Automated verification of source dwell position in lunar ovoid applicators
Journal of Applied Clinical Medical Physics · 2026-11 · 27 (11)
DOI: 10.1002/acm2.70887 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 30.10.2026. Verfasst von Miquel Gubau-Besalú und 6 weiteren Autor:innen. Beteiligte Einrichtung: University of Barcelona, Barcelona, Spain. Hauptthemen laut MEDLINE: Brachytherapy, Software, Radiotherapy Planning, Computer-Assisted, Image Processing, Computer-Assisted. Die Arbeit nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: FSE+.
Abstract (Originalsprache)
Background
Accurate reconstruction of the source dwell positions in high‐dose‐rate (HDR) brachytherapy is essential for treatment quality. The Venezia™ gynaecological applicator, with its lunar‐shaped ovoids, poses challenges because conventional dummy markers do not match the actual source path, introducing systematic uncertainties.
Purpose
The aim of this work was to develop and demonstrate an automated, reproducible, and open‐source software for autoradiographic verification of dwell positions in Venezia lunar ovoids, and to compare measured positions with manufacturer‐provided applicator library data. This pipeline established a standardized approach for clinical commissioning and quality assurance (QA) workflows.
Methods
A custom 3D‐printed holder was designed to standardize autoradiograph acquisition using radiochromic films. An open‐source Python pipeline, referred to as SPICA (Source Position Identification in Circular Applicators), was developed to process scanned films, isolate dwell position, and compute their coordinates with associated uncertainties. The code allows direct comparison between measured positions and manufacturer‐provided applicator library data. We performed preliminary validation of the software by measuring the trajectory of an Ir‐192 source for the 22 and 26 mm ovoids with a step size of 2.5 mm.
Results
Mean discrepancies between measured and reference positions were 1.2 ± 0.5 and 1.5 ± 0.4 mm for the 22 and 26 mm ovoids, respectively, with a maximum deviation of 2.6 mm. Distal positions showed better agreement. These findings confirm that, in the tested applicator‐afterloader combination, vendor libraries provide a generally reliable baseline but may exhibit systematic differences.
Conclusions
The proposed methodology offers a practical, reproducible solution to improve source path verification while reducing manual intervention. It supports more robust commissioning and QA of complex applicators and using the coordinates calculated in the treatment planning software can potentially reduce geometric uncertainties in clinical cases. Individualized commissioning remains essential for applicators with complex geometries.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- University of Barcelona, Barcelona, Spain.
- Computer Vision and Robotics Institute, University of Girona, Girona, Spain.
- Medical Physics Section, Radiation Oncology Department, Hospital Clínic Barcelona, Barcelona, Spain.
- Radiation Protection Department, Hospital Clínic Barcelona, Barcelona, Spain.
- Infrastructure Management, Hospital Clínic Barcelona, Barcelona, Spain.
- Laboratory of Translational Genomics and Targeted therapies in Solid Tumors, Institut d'Investigacions Biomèdiques August Pi i Sunyer, Barcelona, Spain.
ORCID-Profile: Miquel Gubau-Besalú, Adrià Casamitjana, Eduard Agusti-Camprubi, Antonio Herreros
Förderung
- FSE+
Zitieren
APA
Gubau-Besalú, M., Casamitjana, A., Candela-Juan, C., Agusti-Camprubi, E., Segura-Roca, G., Trias, G., & Herreros, A. (2026). Automated verification of source dwell position in lunar ovoid applicators. Journal of Applied Clinical Medical Physics, 27(11). https://doi.org/10.1002/acm2.70887
Vancouver
Gubau-Besalú M, Casamitjana A, Candela-Juan C, Agusti-Camprubi E, Segura-Roca G, Trias G, et al. Automated verification of source dwell position in lunar ovoid applicators. Journal of Applied Clinical Medical Physics. 2026;27(11). doi:10.1002/acm2.70887
BibTeX
@article{GubauBesal2026,
author = {Gubau-Besalú, Miquel and Casamitjana, Adrià and Candela-Juan, Cristian and Agusti-Camprubi, Eduard and Segura-Roca, Gislè and Trias, Gerard and Herreros, Antonio},
title = {Automated verification of source dwell position in lunar ovoid applicators},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {11},
doi = {10.1002/acm2.70887}
}
EUVIMED