Comparison of axial length, intraocular lens power and refractive results using optical and ultrasound biometry
Archives of Ophthalmological Research · 2024
DOI: 10.51271/aor-0012 ↗Kurz gefasst
Fachartikel in der Zeitschrift Archives of Ophthalmological Research, 2024. Verfasst von Ali Ender Kulak und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Aims
The aim of the study was to determine whether a new non-contact partial coherence interferometry method utilized by the Nidek-AL Scan agrees sufficiently with applanation ultrasound A-scan technique in intraocular lens power calculation to replace it.
Methods
This was a prospective study of 150 eyes of 125 patients who underwent cataract surgery with the Nidek AL Scan and ultrasound biometry. The intraocular lens power AL and ACD were compared between two methods with Pearson correlation and Bland-Altman analysis. The mean difference between final spherical equivalent refraction and the prediction with each technology was compared.
Results
On average the axial lengths measured by the Nidek AL-Scan were longer by 0.11 ± 0.13 mm compared the ultrasound biometry. (p0.90, p
Zitieren
APA
Kulak, A. E., & Totan, Y. (2024). Comparison of axial length, intraocular lens power and refractive results using optical and ultrasound biometry. Archives of Ophthalmological Research. https://doi.org/10.51271/aor-0012
Vancouver
Kulak AE, Totan Y. Comparison of axial length, intraocular lens power and refractive results using optical and ultrasound biometry. Archives of Ophthalmological Research. 2024. doi:10.51271/aor-0012
BibTeX
@article{Kulak2024,
author = {Kulak, Ali Ender and Totan, Yüksel},
title = {Comparison of axial length, intraocular lens power and refractive results using optical and ultrasound biometry},
journal = {Archives of Ophthalmological Research},
year = {2024},
doi = {10.51271/aor-0012}
}
EUVIMED