Impact of axial length correction in high intraocular pressure eyes on intraocular lens power calculation
PLOS One · 20.05.2026 · 21 (5) · e0349117
DOI: 10.1371/journal.pone.0349117 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, online erschienen am 20.05.2026. Verfasst von Yuri Kosaka und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Japan Community Health Care Organization, Chukyo Hospital, Nagoya, Japan. Hauptthemen laut MEDLINE: Lenses, Intraocular, Intraocular Pressure, Axial Length, Eye. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This study aimed to investigate the relationship between high intraocular pressure (IOP) effect on axial length (AL) and postoperative refractive error after cataract surgery.
Methods
A retrospective study was conducted on cases that underwent cataract surgery for elevated IOP. AL was corrected based on the change in AL per 1 mmHg of IOP, and the predicted refractive error was evaluated at 3 months postoperatively. Preoperative and postoperative optical biometry was performed using the segmented refractive index method. Intraocular lens (IOL) power was calculated using both the SRK/T and Barrett Universal II (BU II) formulas.
Results
Optical biometry was obtained in 18 cases and 19 eyes before and after cataract surgery. A moderate correlation was observed between the change in IOP and AL (r = 0.687, P = 0.0012), with the change in AL per 1 mmHg of reduction in IOP calculated as 0.005 mm. The mean change in IOP was 38.87 ± 10.80 mmHg (P = 0.001), and the AL \ significantly shortened by 0.18 ± 0.09 mm (P = 0.001). The mean postoperative refractive errors were 0.36 ± 0.59 D with the SRK/T formula, 0.79 ± 0.58 D with the BU II formula, showing significantly more hyperopia (p = 0.001). After AL correction, the predicted refractive error was reduced to −0.08 ± 0.54 D (SRK/T) and 0.27 ± 0.55 D (BU II).
Conclusions
In cataract surgery for eyes with elevated IOP, using corrected AL based on the anticipated IOP reduction appears to decrease the predicted refractive error. Additionally, it should be noted that the BU II formula may result in a hyperopic shift in refractive prediction error, even when using the adjusted AL.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Ophthalmology, Japan Community Health Care Organization, Chukyo Hospital, Nagoya, Japan.
- Nagoya Eye Clinic, Nagoya, Japan.
- Major of Vision Sciences, Faculty of Health & Medical Sciences, Aichi Shukutoku University, Aichi, Japan.
- Major of Vision Sciences, Graduate School of Psychology and Medical Sciences, Aichi Shukutoku University, Aichi, Japan.
ORCID-Profile: Takashi Kojima, Akeno Tamaoki
Zitieren
APA
Kosaka, Y., Kojima, T., Tamaoki, A., Takagi, Y., Sawaki, A., Nogami, T., & Kaga, T. (2026). Impact of axial length correction in high intraocular pressure eyes on intraocular lens power calculation. PLOS One, 21(5), e0349117. https://doi.org/10.1371/journal.pone.0349117
Vancouver
Kosaka Y, Kojima T, Tamaoki A, Takagi Y, Sawaki A, Nogami T, et al. Impact of axial length correction in high intraocular pressure eyes on intraocular lens power calculation. PLOS One. 2026;21(5):e0349117. doi:10.1371/journal.pone.0349117
BibTeX
@article{Kosaka2026,
author = {Kosaka, Yuri and Kojima, Takashi and Tamaoki, Akeno and Takagi, Yuki and Sawaki, Ayako and Nogami, Tsuyoshi and Kaga, Tatsushi},
title = {Impact of axial length correction in high intraocular pressure eyes on intraocular lens power calculation},
journal = {PLOS One},
year = {2026},
volume = {21},
number = {5},
pages = {e0349117},
doi = {10.1371/journal.pone.0349117}
}
EUVIMED