EUVIMEDEUVIMED entdecken
QuerschnittstudieOpen Access · CC BY-NC-ND 4.0

Comparison of peek acuity mobile application with the standard snellen’s chart for recording visual acuity: a cross-sectional study

Elakkiya, VENKATESAN, Dr. Manjusree

SJ Ophthalmology Africa · 30.10.2026 · 3 (4) · 9

DOI: 10.51168/b4g3kd62 ↗

Kurz gefasst

Querschnittstudie in der Zeitschrift SJ Ophthalmology Africa, online erschienen am 30.10.2026. Verfasst von Elakkiya und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Visual acuity (VA) assessment is the cornerstone of ophthalmic examination. While the Snellen’s chart remains the clinical standard, its lack of portability and requirement for standardised lighting limit its utility in outreach settings. The Peek Acuity mobile application offers a portable, standardised digital alternative.1 Objectives: To evaluate the clinical validity and reliability of the Peek Acuity mobile application compared to the Snellen’s chart in an adult population.

Methods

This hospital-based, cross-sectional study in Tamil Nadu enrolled 100 participants (200 eyes) aged 18–88 years (mean 42.2 ± 16.2 years). Inclusion required VA ≥ 6/60. Participants were tested using the Peek Acuity app (2 metres, 100% brightness) and a retro-illuminated Snellen’s chart (6 metres). Snellen results were converted to logMAR for analysis. Statistical measures included Pearson’s correlation (r) and Bland–Altman plots for limits of agreement (LoA).

Results

Analysis of 199 valid data points showed a mean logMAR VA of 0.283 ± 0.255 (Snellen) and 0.261 ± 0.253 (Peek). A very strong positive correlation was observed (r = 0.970, P < 0.001). Bland–Altman analysis revealed a clinically negligible mean bias of 0.022 logMAR. The 95% LoA ranged from −0.100 to +0.144 logMAR, indicating that 95% of measurements fell within 1.5 lines of the standard. Eye-wise means were comparable (right eye 0.279 vs 0.256; left eye 0.286 vs 0.266); both methods shared a median of 0.18 logMAR (SD of differences 0.062).

Conclusion

Peek Acuity demonstrates excellent clinical validity and strong agreement with the Snellen’s chart. Its portability and built-in standardisation make it a reliable, interchangeable tool for vision screening and triage, particularly in resource-limited or confined clinical environments. Recommendations: Peek Acuity can be adopted for adult vision screening in outreach camps and space-limited clinics, with confirmatory Snellen testing for borderline results. Validation in low-vision, paediatric and multi-device settings is recommended.

Zitieren

APA

Elakkiya, VENKATESAN, & Manjusree, D. (2026). Comparison of peek acuity mobile application with the standard snellen’s chart for recording visual acuity: a cross-sectional study. SJ Ophthalmology Africa, 3(4), 9. https://doi.org/10.51168/b4g3kd62

Vancouver

Elakkiya, VENKATESAN, Manjusree D. Comparison of peek acuity mobile application with the standard snellen’s chart for recording visual acuity: a cross-sectional study. SJ Ophthalmology Africa. 2026;3(4):9. doi:10.51168/b4g3kd62

BibTeX
@article{Elakkiya2026,
  author = {Elakkiya and VENKATESAN and Manjusree, Dr.},
  title = {Comparison of peek acuity mobile application with the standard snellen’s chart for recording visual acuity: a cross-sectional study},
  journal = {SJ Ophthalmology Africa},
  year = {2026},
  volume = {3},
  number = {4},
  pages = {9},
  doi = {10.51168/b4g3kd62}
}
Kostenlos starten →