Consequences of Pausing Heated Humidification During Invasive Ventilation

François Lellouche, Pierre-Alexandre Bouchard

Respiratory Care · 2024

DOI: 10.4187/respcare.12084 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Respiratory Care, 2024. Verfasst von François Lellouche und 1 weiteren Autor:innen. Beteiligte Einrichtung: Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Québec, Canada. francois.lellou…. Hauptthemen laut MEDLINE: Humidity, Humidifiers, Respiration, Artificial, Hot Temperature. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: During invasive mechanical ventilation, where medical gases are very dry and the upper airway is bypassed, appropriate gas conditioning and humidification are mandatory at all times. Results of in vitro studies suggest that dry gases may improve lung deposition during nebulization, but this has not been confirmed through in vivo studies. The objective of this study was to measure gas humidity under multiple conditions to better describe gas hygrometry when heated humidifiers are turned off. METHODS: We measured, on a bench, the hygrometry of different gases at steady state: medical gases, at the Y-piece without humidifier, with the humidifier switched off, and with humidifier switched on. We measured gas humidity every 10–60 s during dynamic conditions after switching off the heated humidifier and after switching on the heated humidifier. Hygrometry was measured by using the psychrometric method with at least 3 measurements for each tested condition. RESULTS: We performed 287 psychrometric measurements in different situations. The mean ± SD gas absolute humidity at steady state during different conditions were the following: 1.6 ± 0.2 mg H 2 O/L for the medical gases, 4.5 ± 0.9 mg H 2 O/L at the Y-piece without humidifier, 9.1 ± 0.3 mg H 2 O/L at the Y-piece with heated humidifier turned off, and 34.2 ± 2.2 mg H 2 O/L at the Y-piece with the heated humidifier turned on. During the dynamic evaluation, after turning off the humidifier, humidity was < 30 mg H 2 O/L after a few minutes, attained 15 mg H 2 O/L after 15 min, and was below 10 mg H 2 O/L after 1 h but never reached the level of dry medical gases. After turning on the heated humidifier, the gas hygrometry reached 30 mg H 2 O/L after 5 min. CONCLUSIONS: When heated humidifiers are turned off, gas humidity levels are very low but not as low as medical gases. The clinical impact of repeated shutdowns is unknown. As recommended, heated humidifiers should never be turned off during nebulization.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HumidityHumidifiersRespiration, ArtificialHot TemperatureNebulizers and VaporizersHumansGasesEquipment Design

Stichwörter der Autor:innen

absolute humidityendotracheal tube occlusionsheated humidifiernebulizationpsychrometry

Substanzen

Gases

Autor:innen und Institutionen

  • Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Québec, Canada. francois.lellouche@criucpq.ulaval.ca.
  • Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Québec, Canada.

Zitieren

APA

Lellouche, F., & Bouchard, P. A. (2024). Consequences of Pausing Heated Humidification During Invasive Ventilation. Respiratory Care. https://doi.org/10.4187/respcare.12084

Vancouver

Lellouche F, Bouchard PA. Consequences of Pausing Heated Humidification During Invasive Ventilation. Respiratory Care. 2024. doi:10.4187/respcare.12084

BibTeX
@article{Lellouche2024,
  author = {Lellouche, François and Bouchard, Pierre-Alexandre},
  title = {Consequences of Pausing Heated Humidification During Invasive Ventilation},
  journal = {Respiratory Care},
  year = {2024},
  doi = {10.4187/respcare.12084}
}
Kostenlos starten →