Evaluation of the Humidification Performances of New Generation of Heated Wire Humidifiers
Respiratory Care · 2024
DOI: 10.4187/respcare.11352 ↗Kurz gefasst
Fachartikel in der Zeitschrift Respiratory Care, 2024. Verfasst von François Lellouche und 2 weiteren Autor:innen. Beteiligte Einrichtung: Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Quebec (Quebec), Canada. francois.lellouche@criucpq.ulava…. Hauptthemen laut MEDLINE: Humidifiers, Respiration, Artificial. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
BACKGROUND: At high ambient temperatures in ICU rooms, the humidification performances of heated-wire humidifiers are significantly reduced, with delivered gas humidity well below 30 mg H 2 O/L, which leads to an increased risk of endotracheal occlusions, subocclusions, or mucociliary dysfunction. The objective of the study was to evaluate the humidity delivered at the Y-piece with new-generation heated-wire humidifiers with advanced algorithm (FP950 [Fisher & Paykel Healthcare, Auckland, New Zealand] and VHB20 [Vincent Medical, Inspired, Hong Kong]) while varying ambient temperatures. METHODS: We measured, on the bench, the hygrometry of inspiratory gases delivered by a new generation of heated-wire humidifiers (i) FP950, (ii) VHB20 and a previous generation of heated-wire humidifiers, (iii) MR850 (Fisher & Paykel) with the usual settings (37°C at the chamber/40°C at the Y-piece), (iv) MR850 with no temperature gradient (40°C/40°C), and (v) MR850 with the automatic compensation algorithm activated. Hygrometry was measured with the psychrometric method after 1 h of stability while varying the room temperature from 20 to 30°C. RESULTS: Two hundred ninety-four hygrometric bench measurements were performed at steady state for the different tested conditions. With the new heated-wire humidifiers (FP950 and VHB20), gas humidity delivered remained > 30 mg H 2 O/L in all tested conditions, even at high ambient temperatures (>25°C). With previous generations of heated-wire humidifiers (MR850), at high ambient temperature, humidity delivered was adequate in only 26% (11/42) of the measurements when the usual settings were used (37°C/40°C) and 30% (11/37) with automatic compensation. When no temperature gradient was set (40°C/40°C), humidity delivered was > 30 mg H 2 O/L in 91% (30/33) of the measurements at a high ambient temperature. With an ambient temperature < 25°C, almost all devices and settings provided adequate humidity. CONCLUSIONS: The new FP950 and VHB20 heated-wire humidifiers by using advanced algorithms demonstrated stable performance while varying the ambient temperature by 20–30°C, better than the previous generation of heated humidifiers when ambient temperatures were high.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Quebec (Quebec), Canada. francois.lellouche@criucpq.ulaval.ca.
- Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Quebec (Quebec), Canada.
Zitieren
APA
Lellouche, F., Rousseau, E., & Bouchard, P. A. (2024). Evaluation of the Humidification Performances of New Generation of Heated Wire Humidifiers. Respiratory Care. https://doi.org/10.4187/respcare.11352
Vancouver
Lellouche F, Rousseau E, Bouchard PA. Evaluation of the Humidification Performances of New Generation of Heated Wire Humidifiers. Respiratory Care. 2024. doi:10.4187/respcare.11352
BibTeX
@article{Lellouche2024,
author = {Lellouche, François and Rousseau, Emilie and Bouchard, Pierre-Alexandre},
title = {Evaluation of the Humidification Performances of New Generation of Heated Wire Humidifiers},
journal = {Respiratory Care},
year = {2024},
doi = {10.4187/respcare.11352}
}
EUVIMED