Temperature and Humidity During Different Neonatal Respiratory Support Modes: An In‐Vitro Performance of Four Heated Humidifiers

Emanuela Zannin iD, Camilla Rigotti iD, Stefano Robbiani, Raffale L Dellacà iD, Maria Luisa Ventura

Pediatric Pulmonology · 2025

DOI: 10.1002/ppul.71184 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Pediatric Pulmonology, 2025. Verfasst von Emanuela Zannin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Neonatal Intensive Care Unit, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy. Hauptthemen laut MEDLINE: Humidity, Humidifiers, Temperature, High-Frequency Ventilation. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Adequate heating and humidification of inspired gas are crucial in preterm infants requiring respiratory support. Our objective was to compare the temperature (T) and absolute humidity (AH) achieved with different neonatal heated humidifiers commonly used during different respiratory support modes and respiratory settings.

Methods

Four neonatal heated humidifiers were tested at different gas flow rates and during conventional and high‐frequency oscillatory ventilation at clinically relevant settings. The patient was simulated using a neonatal test lung placed in a neonatal incubator warmed at 37°C. Temperature and relative humidity were measured using a thermohygrometer placed proximal to the test lung and used to calculate AH.

Results

Measured temperature and humidity varied among heated humidifiers and respiratory support settings. The bias flow was the primary determinant of the measured temperature and humidity: temperature and AH significantly increased with flow. When the flow through the chamber was below 5 L/min the MR950 produced T < 36.7°C and AH < 37 g/m 3 proximal to the patient. The Hydraltis 9500neo with chamber and proximal temperature set to 38°C and 40°C and the MR850 produced AH > 44 g/m 3 in most conditions.

Conclusion

The temperature and humidity proximal to the patients' airways might vary considerably based on the heated humidifiers and the flow through the chamber. Further clinical studies are needed to evaluate the effect of such variability on the patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HumidityHumidifiersTemperatureHigh-Frequency VentilationRespiration, ArtificialHumansInfant, NewbornInfant, PrematureHot Temperature

Stichwörter der Autor:innen

humidifiershumidityinfantmechanical ventilationprematuretemperature

Autor:innen und Institutionen

  • Neonatal Intensive Care Unit, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy.
  • TechRes Lab, Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milano, Italy.

ORCID-Profile: Emanuela Zannin, Camilla Rigotti, Raffale L Dellacà

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Zannin, E., Rigotti, C., Robbiani, S., Dellacà, R. L., & Ventura, M. L. (2025). Temperature and Humidity During Different Neonatal Respiratory Support Modes: An In‐Vitro Performance of Four Heated Humidifiers. Pediatric Pulmonology. https://doi.org/10.1002/ppul.71184

Vancouver

Zannin E, Rigotti C, Robbiani S, Dellacà RL, Ventura ML. Temperature and Humidity During Different Neonatal Respiratory Support Modes: An In‐Vitro Performance of Four Heated Humidifiers. Pediatric Pulmonology. 2025. doi:10.1002/ppul.71184

BibTeX
@article{Zannin2025,
  author = {Zannin, Emanuela and Rigotti, Camilla and Robbiani, Stefano and Dellacà, Raffale L and Ventura, Maria Luisa},
  title = {Temperature and Humidity During Different Neonatal Respiratory Support Modes: An In‐Vitro Performance of Four Heated Humidifiers},
  journal = {Pediatric Pulmonology},
  year = {2025},
  doi = {10.1002/ppul.71184}
}
Kostenlos starten →