External Jet Nebulization and Measured Ventilator Performance
Respiratory Care · 2024
DOI: 10.4187/respcare.11296 ↗Kurz gefasst
Fachartikel in der Zeitschrift Respiratory Care, 2024. Verfasst von Jeyanthan Jayakumaran und 2 weiteren Autor:innen. Beteiligte Einrichtung: The Division of Pulmonary, Critical Care and Sleep Medicine, Department of Medicine, Stony Brook University Medical Center, Stony Brook, New York. Hauptthemen laut MEDLINE: Ventilators, Mechanical, Tidal Volume, Nebulizers and Vaporizers. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
BACKGROUND: During invasive ventilation, external flow jet nebulization results in increases in displayed exhaled tidal volumes (VT). We hypothesized that the magnitude of the increase is inaccurate. An ASL 5000 simulator measured ventilatory parameters over a wide range of adult settings: actual VT, peak inspiratory pressure (PIP), and time to minimum pressure. METHODS: Ventilators with internal and external flow sensors were tested by using a variety of volume and pressure control modes (the target VT was 420 mL). Patient conditions (normal, COPD, ARDS) defined on the ASL 5000 were assessed at baseline and with 3.5 or 8 L/min of added external flow. Patient-triggering was assessed by reducing muscle effort to the level that resulted in backup ventilation and by changing ventilator sensitivity to the point of auto-triggering. RESULTS: Results are reported as percentage change from baseline after addition of 3.5 or 8 L/min external flow. For ventilators with internal flow sensors, changes in displayed exhaled VT ranged from 10% to 118%, however, when using volume control, actual increases in actual VT and PIP were only 4%-21% (P = .063, .031) and 6%-24% (P = .25, .031), respectively. Changes in actual VT correlated closely with changes in PIP (P < .001; R2 = 0.68). For pressure control, actual VT decreased by 3%-5% (P = .031) and 4%-9% (P = .031) with 3.5 and 8 L/min respectively, PIP was unchanged. With external flow sensors at the distal Y-piece junction, volume and pressure changes were statistically insignificant. The time to minimum pressure increased at most by 8% (P = .02) across all modes and ventilators. The effects on muscle pressure were minimal (∼1 cm H2O), and ventilator sensitivity effects were nearly undetectable. CONCLUSIONS: External flow jet nebulization resulted in much smaller changes in volume than indicated by the ventilator display. Statistically significant effects were confined primarily to machines with internal flow sensors. Differences approached the manufacturer-reported variation in ventilator baseline performance. During nebulizer therapy, effects on VT can be estimated at the bedside by monitoring PIP.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The Division of Pulmonary, Critical Care and Sleep Medicine, Department of Medicine, Stony Brook University Medical Center, Stony Brook, New York.
- The Respiratory Care Program, School of Health Professions, Stony Brook University, Stony Brook, New York ann.cuccia@stonybrook.edu.
Zitieren
APA
Jayakumaran, J., Smaldone, G. C., & Cuccia, A. D. (2024). External Jet Nebulization and Measured Ventilator Performance. Respiratory Care. https://doi.org/10.4187/respcare.11296
Vancouver
Jayakumaran J, Smaldone GC, Cuccia AD. External Jet Nebulization and Measured Ventilator Performance. Respiratory Care. 2024. doi:10.4187/respcare.11296
BibTeX
@article{Jayakumaran2024,
author = {Jayakumaran, Jeyanthan and Smaldone, Gerald C and Cuccia, Ann D},
title = {External Jet Nebulization and Measured Ventilator Performance},
journal = {Respiratory Care},
year = {2024},
doi = {10.4187/respcare.11296}
}
EUVIMED