The in-vitro performance of invasive mechanical ventilators in lung model
Science Progress · 2026-04 · 109 (2)
DOI: 10.1177/00368504261444846 ↗Kurz gefasst
Fachartikel in der Zeitschrift Science Progress, online erschienen am 27.04.2026. Verfasst von Xingshuo Hu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Respiratory and Critical Care Medicine, Tianjin Medical University General Hospital, Tianjin, China. Hauptthemen laut MEDLINE: Ventilators, Mechanical, Lung, Models, Biological. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Objective
To evaluate the technical performance and patient-ventilator synchronization of different invasive mechanical ventilators.
Methods
Six ventilators were tested using the ASL5000 active lung simulator. Ventilators were categorized by air supply type: central (Dräger V600, Maquet Servo-u, Mindray SV800) and turbine-based (Savina 300, SV300, Resvent RS300). The measured parameters included tidal volume (VT), plateau pressure (Pplat), positive end-expiratory pressure (PEEP), peak pressure (Ppeak), time to reach 90% of inspiratory pressure (T (90%)), triggering delay (DT), and leak compensation. Ventilators were evaluated in volume assist-control (V-A/C), pressure assist-control (P-A/C), and pressure support ventilation (PSV) modes under acute respiratory distress syndrome (ARDS) and chronic obstructive pulmonary disease (COPD) conditions using ASL5000™ lung simulator, with or without intentional leak (4.4 L/min). For each ventilator under these different conditions, we analyzed pressurization accuracy, VT accuracy, inspiratory muscle effort, and response efficiency.
Results
The pressurization accuracy of ventilators with central air supply did not differ significantly from that of turbine-driven ventilators during V-A/C and P-A/C ventilation, irrespective of spontaneous breathing or air leakage. The mean absolute errors of target VT were 8.9% (V600), 10.3% (Servo-U), 14.4% (SV800), 4.6% (Savina300), 18.8% (SV300), and 19.9% (RS300), and ventilators with central air supply generally have better VT accuracy, except for Savina300 ( P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Respiratory and Critical Care Medicine, Tianjin Medical University General Hospital, Tianjin, China.
- Critical Care Unit Within the Department of Respiratory and Critical Care Medicine, 8th Medical Center of Chinese PLA General Hospital, Beijing, China.
- College of Pulmonary and Critical Care Medicine, 8th Medical Center of Chinese PLA General Hospital, Beijing, China.
- Haihe Clinical School, Tianjin Medical University, Tianjin, China.
ORCID-Profile: Qi Wu
Zitieren
APA
Hu, X., Ma, Y., Yang, Q., Xie, L., & Wu, Q. (2026). The in-vitro performance of invasive mechanical ventilators in lung model. Science Progress, 109(2). https://doi.org/10.1177/00368504261444846
Vancouver
Hu X, Ma Y, Yang Q, Xie L, Wu Q. The in-vitro performance of invasive mechanical ventilators in lung model. Science Progress. 2026;109(2). doi:10.1177/00368504261444846
BibTeX
@article{Hu2026,
author = {Hu, Xingshuo and Ma, Yizhou and Yang, Qingyun and Xie, Lixin and Wu, Qi},
title = {The in-vitro performance of invasive mechanical ventilators in lung model},
journal = {Science Progress},
year = {2026},
volume = {109},
number = {2},
doi = {10.1177/00368504261444846}
}
EUVIMED