To determine the effect of plasma nitriding treatment 56 on screw loosening and surface topography of different 78 implant–abutment screw systems with and without thermocycling: An in vitro study
The Journal of Indian Prosthodontic Society · 2023 · 23 (3) · 285-293
DOI: 10.4103/jips.jips_147_23 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Indian Prosthodontic Society, online erschienen am 18.07.2023. Verfasst von Nengpichong Haokip und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics and Crown and Bridge, School of Dental Sciences, Krishna Vishwa Vidyapeeth Deemed to be University, Karad, Maharashtra, …. Hauptthemen laut MEDLINE: Dental Implants. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Aim
The aim of this study was to evaluate and compare the effect of plasma nitride-treated abutment screws of two different implant systems on screw loosening and surface topography with and without thermocycling. Settings and Design: This was an in-vitro experimental study. Materials and Methods: Fifty-two abutment screws (Group A: 26 Genesis and Group B: 26 Bredent) underwent plasma nitride treatment and were subdivided into two groups, one without thermocycling and one with thermocycling. Dynamic load was applied and detorque values were evaluated for determining the screw loosening using “independent t-test” with the help of IBM SPSS Statistics 20 and scanning electron microscopy was done to check for surface topography. Statistical Analysis Used: Inter- and intragroup comparisons were done using independent t-test (SPSS: Statistical Package for the Social Sciences software version 20).
Results
Plasma nitriding treatment genesis implant system abutment screw showed more screw loosening ( P < 0.05) and surface roughness as compared to bredent with and without thermocycling.
Conclusion
From the present study, it was shown that plasma nitride-treated abutment screws decreased the occurrence of screw loosening favoring the bredent implant–abutment system more than the genesis implant–abutment system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Prosthodontics and Crown and Bridge, School of Dental Sciences, Krishna Vishwa Vidyapeeth Deemed to be University, Karad, Maharashtra, India.
Zitieren
APA
Haokip, N., Tewary, S., Sanyal, P., & Kamnoor, S. (2023). To determine the effect of plasma nitriding treatment 56 on screw loosening and surface topography of different 78 implant–abutment screw systems with and without thermocycling: An in vitro study. The Journal of Indian Prosthodontic Society, 23(3), 285-293. https://doi.org/10.4103/jips.jips_147_23
Vancouver
Haokip N, Tewary S, Sanyal P, Kamnoor S. To determine the effect of plasma nitriding treatment 56 on screw loosening and surface topography of different 78 implant–abutment screw systems with and without thermocycling: An in vitro study. The Journal of Indian Prosthodontic Society. 2023;23(3):285-293. doi:10.4103/jips.jips_147_23
BibTeX
@article{Haokip2023,
author = {Haokip, Nengpichong and Tewary, Shivsagar and Sanyal, Pronob and Kamnoor, Shubha},
title = {To determine the effect of plasma nitriding treatment 56 on screw loosening and surface topography of different 78 implant–abutment screw systems with and without thermocycling: An in vitro study},
journal = {The Journal of Indian Prosthodontic Society},
year = {2023},
volume = {23},
number = {3},
pages = {285--293},
doi = {10.4103/jips.jips_147_23}
}
EUVIMED