Assessing the analytical performance of the routine chemistry analytes in the clinical biochemistry laboratory by application of Six Sigma metrics
Journal of Clinical and Scientific Research · 2026-01 · 15 (1) · 38-46
DOI: 10.4103/jcsr.jcsr_28_25 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical and Scientific Research, online erschienen am 07.05.2026. Verfasst von Mamatha Kathula und 2 weiteren Autor:innen. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background: Clinical biochemistry laboratory plays a critical role in patient management. Hence, the laboratory must report accurate and reliable results through proper implementation of quality control (QC) practices. This study was conducted to evaluate the analytical performance of routine chemistry analytes in our tertiary care hospital laboratory by the application of Sigma metrics to determine the underlying causes of unsatisfactory performance of analytes with a Sigma value of
Zitieren
APA
Kathula, M., Kiramayi, V. S., & Bitla, A. R. (2026). Assessing the analytical performance of the routine chemistry analytes in the clinical biochemistry laboratory by application of Six Sigma metrics. Journal of Clinical and Scientific Research, 15(1), 38-46. https://doi.org/10.4103/jcsr.jcsr_28_25
Vancouver
Kathula M, Kiramayi VS, Bitla AR. Assessing the analytical performance of the routine chemistry analytes in the clinical biochemistry laboratory by application of Six Sigma metrics. Journal of Clinical and Scientific Research. 2026;15(1):38-46. doi:10.4103/jcsr.jcsr_28_25
BibTeX
@article{Kathula2026,
author = {Kathula, Mamatha and Kiramayi, V. S. and Bitla, Aparna R.},
title = {Assessing the analytical performance of the routine chemistry analytes in the clinical biochemistry laboratory by application of Six Sigma metrics},
journal = {Journal of Clinical and Scientific Research},