Reliability and Validity of the Martin Vigorimeter for Grip Strength Measurement in Korean Adults
Clinics in Orthopedic Surgery · 2024
DOI: 10.4055/cios23383 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Clinics in Orthopedic Surgery, 2024. Verfasst von Jeong Kil Lee und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Chungnam National University Sejong Hospital, Chungnam National University College of Medicine, Sejong, Korea. Hauptthemen laut MEDLINE: Hand Strength, Muscle Strength Dynamometer. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Chungnam National University Hospital.
Abstract (Originalsprache)
Background
Grip strength is important for fine motor skills, and one of the measurement tools for grip strength is the Martin Vigorimeter (MV) dynamometer. Studies on establishing the reliability and validity of the MV in Koreans are limited. We aimed to establish the reliability and validity of the MV for grip strength measurement in healthy Korean adults by comparing it with the Jamar dynamometer, the standard tool used by the American Society of Hand Therapists.
Methods
In total, 99 healthy participants (50 men and 49 women) were enrolled. Grip strength was measured using the Jamar dynamometer and MV. Reliability and validity were assessed using the intraclass correlation coefficient (ICC) and minimal detectable change (MDC). The correlation between the measurements of the instruments was analyzed using Pearson's correlation. The effect of hand anthropometry was evaluated, and the conversion equation between the instruments was calculated.
Results
MV showed excellent reliability (ICC > 0.90, p < 0.001) and validity with a high correlation (0.7 ≤ r < 0.9) with the Jamar dynamometer. The MDC was acceptable for detecting minimal clinically important differences (< 19.5%) in both instruments (Jamar: 3.4%-6.7%, MV: 3.8% to 6.3%). The grip strength measured using the MV was independent of hand anthropometry, unlike that using the Jamar dynamometer.
Conclusions
This study provides insights into the relationship between the Jamar and MV instruments for measuring grip strength in Koreans. The MV is a viable alternative to the Jamar dynamometer in Koreans, offering not only reproducible and reliable measurements of grip strength but also the advantage of being unaffected by variations in hand anthropometry.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedic Surgery, Chungnam National University Sejong Hospital, Chungnam National University College of Medicine, Sejong, Korea.
- Department of Biomedical Engineering, Chungnam National University, Sejong, Korea.
- Research Institute for Future Medical Science, Chungnam National University Sejong Hospital, Sejong, Korea.
ORCID-Profile: Jeong Kil Lee, Meesun Jung, Han Byul Lee, Hyung Jin Chung, Seung Hoo Lee
Förderung
- Chungnam National University Hospital
Zitieren
APA
Lee, J. K., Jung, M., Lee, H. B., Chung, H. J., & Lee, S. H. (2024). Reliability and Validity of the Martin Vigorimeter for Grip Strength Measurement in Korean Adults. Clinics in Orthopedic Surgery. https://doi.org/10.4055/cios23383
Vancouver
Lee JK, Jung M, Lee HB, Chung HJ, Lee SH. Reliability and Validity of the Martin Vigorimeter for Grip Strength Measurement in Korean Adults. Clinics in Orthopedic Surgery. 2024. doi:10.4055/cios23383
BibTeX
@article{Lee2024,
author = {Lee, Jeong Kil and Jung, Meesun and Lee, Han Byul and Chung, Hyung Jin and Lee, Seung Hoo},
title = {Reliability and Validity of the Martin Vigorimeter for Grip Strength Measurement in Korean Adults},
journal = {Clinics in Orthopedic Surgery},
year = {2024},
doi = {10.4055/cios23383}
}
EUVIMED