Optimization of Surgical Drill Margin Dimension to Reduce Bone Temperature
Key Engineering Materials · 04.05.2026 · 1053 · 27-37
DOI: 10.4028/p-tq4lrt ↗Kurz gefasst
Fachartikel in der Zeitschrift Key Engineering Materials, online erschienen am 04.05.2026. Verfasst von Mohd Faizal Ali Akhbar und 2 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben.
Abstract (Originalsprache)
Excessive heat generation during bone drilling is a leading cause of thermal osteonecrosis—a serious risk in medical departments. Despite extensive drill design research, the influence of margin geometry remains underexplored. This study presents finite element modeling and statistical optimization to evaluate and optimize drill margin geometry—specifically margin width ( Mw ) and height ( Mh )—to reduce bone temperature rise during surgery. A thermo-mechanical finite element model was developed in DEFORM-3D to simulate cortical bone drilling using drill bits with varied margin dimensions. The models were validated experimentally using bovine cortical bone, with an average temperature prediction errors of 2.4–8.0%. The maximum bone temperature ( T max ) was selected as the objective function. A central composite design (CCD) was applied to generate experimental runs, followed by response surface methodology (RSM) and desirability-based optimization. The second-order effect of Mw contributed 47.2% to T max . The optimal Mh (0.05 mm) and Mw (0.22 mm)—with a desirability value of 0.985—could reduce T max below the osteonecrosis level with only a 44.8 °C temperature rise. This study demonstrates a novel computational approach for optimizing surgical drill margins—a previously underutilized parameter. The findings may support future developments in drill bit customization and robotic surgery systems to minimize thermal injury to bone cells.
Zitieren
APA
Akhbar, M. F. A., Jamaludin, S., & Alias, R. (2026). Optimization of Surgical Drill Margin Dimension to Reduce Bone Temperature. Key Engineering Materials, 1053, 27-37. https://doi.org/10.4028/p-tq4lrt
Vancouver
Akhbar MFA, Jamaludin S, Alias R. Optimization of Surgical Drill Margin Dimension to Reduce Bone Temperature. Key Engineering Materials. 2026;1053:27-37. doi:10.4028/p-tq4lrt
BibTeX
@article{Akhbar2026,
author = {Akhbar, Mohd Faizal Ali and Jamaludin, Shahrizan and Alias, Rodianah},
title = {Optimization of Surgical Drill Margin Dimension to Reduce Bone Temperature},
journal = {Key Engineering Materials},
year = {2026},
volume = {1053},
pages = {27--37},
doi = {10.4028/p-tq4lrt}
}
EUVIMED