EUVIMEDEUVIMED entdecken
Crossref

Thermal performance of modified surgical drill margins during bone drilling: An experimental, numerical, and optimization case study

Mohd Faizal Ali Akhbar iD, Shahrizan Jamaludin, Suriani Mat Jusoh, Fatin Alias, Mohd Azlan Musa, Rodianah Alias

Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · 2026-04 · 240 (4) · 411-426

DOI: 10.1177/09544119261426440 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, online erschienen am 07.03.2026. Verfasst von Mohd Faizal Ali Akhbar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Naval Architecture and Maritime Technology, Faculty of Ocean Engineering Technology (FTKK), Universiti Malaysia Terengganu (UMT), Teren…. Hauptthemen laut MEDLINE: Temperature, Bone and Bones, Orthopedic Procedures, Mechanical Phenomena. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 39 Literaturangaben. Gefördert von: Ministry of Higher Education, Malaysia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Excessive heat generation during bone drilling causes thermal osteonecrosis, posing a significant risk in orthopedic and dental procedures. While various aspects of drill design have been studied, the specific influence of margin geometry remains underexplored in the context of thermal engineering. This study integrates drilling simulation, experimental validation, and statistical optimization to evaluate the thermal impact of drill margin width ( M w ) and height ( M h ) during cortical bone drilling. A validated thermo-mechanical model was developed using commercial software DEFORM-3D. The simulation results were validated with experimental bone drilling with small temperature prediction errors (2.4% and 8.0%). Key thermal metrics (maximum temperature ( T max ), osteonecrosis diameter (OD), and osteonecrosis depth (OH)) were optimized using a central composite design (CCD) of response surface methodology (RSM) and desirability-based multi-objective optimization. Results revealed that M w had the most significant second-order influence on T max (47.2%), while M h dominated OD (41.1%) and OH (47.8%). The optimal drill margins ( M h = 0.05 mm and M w = 0.22 mm), which achieved a desirability score of 0.985, could reduce T max by up to 44.8°C, which is below the osteonecrosis threshold. This work highlights drill margins as a critical yet previously underutilized design variable, offering an alternative pathway for the thermally optimized development of surgical tools and next-generation robotic-assisted drilling systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TemperatureBone and BonesOrthopedic ProceduresMechanical PhenomenaEquipment DesignOsteonecrosisAnimalsFinite Element AnalysisComputer Simulation

Stichwörter der Autor:innen

bonedrillingfinite element modelingmargin heightmargin widthmulti-objective optimizationthermal osteonecrosis

Autor:innen und Institutionen

  • Department of Naval Architecture and Maritime Technology, Faculty of Ocean Engineering Technology (FTKK), Universiti Malaysia Terengganu (UMT), Terengganu, Malaysia.
  • Faculty of Innovative Design and Technology, Universiti Sultan Zainal Abidin, Kuala Terengganu, Terengganu, Malaysia.

ORCID-Profile: Mohd Faizal Ali Akhbar

Förderung

  • Ministry of Higher Education, Malaysia

Zitieren

APA

Akhbar, M. F. A., Jamaludin, S., Mat Jusoh, S., Alias, F., Musa, M. A., & Alias, R. (2026). Thermal performance of modified surgical drill margins during bone drilling: An experimental, numerical, and optimization case study. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 240(4), 411-426. https://doi.org/10.1177/09544119261426440

Vancouver

Akhbar MFA, Jamaludin S, Mat Jusoh S, Alias F, Musa MA, Alias R. Thermal performance of modified surgical drill margins during bone drilling: An experimental, numerical, and optimization case study. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2026;240(4):411-426. doi:10.1177/09544119261426440

BibTeX
@article{Akhbar2026,
  author = {Akhbar, Mohd Faizal Ali and Jamaludin, Shahrizan and Mat Jusoh, Suriani and Alias, Fatin and Musa, Mohd Azlan and Alias, Rodianah},
  title = {Thermal performance of modified surgical drill margins during bone drilling: An experimental, numerical, and optimization case study},
  journal = {Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine},
  year = {2026},
  volume = {240},
  number = {4},
  pages = {411--426},
  doi = {10.1177/09544119261426440}
}
Kostenlos starten →