Prolonged Rewarming Was Associated with Reduced Alkaline Phosphatase Bias in External Quality Assessment Materials: A Laboratory Study

Dayun Xie, Xue Wen, Baoxiu Chen, Yali Li, Man Zhu, Wenbiao Zhou, Jiao Liu

Journal of Visualized Experiments · 03.11.2026 · (237)

DOI: 10.3791/75749 ↗

Kurz gefasst

Video/Audio in der Zeitschrift Journal of Visualized Experiments, online erschienen am 03.11.2026. Verfasst von Dayun Xie und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Chuxiong Yi Autonomous Prefecture People's Hospital. Hauptthemen laut MEDLINE: Alkaline Phosphatase, Specimen Handling.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Recurrent negative alkaline phosphatase (ALP) bias in external quality assessment (EQA) materials despite stable internal quality control prompted investigation of handling conditions. This study examined elapsed time after removal from frozen storage in five independent EQA materials. Materials were separately aliquoted (200 µL) and stored at -20 °C before testing. Different aliquots were analyzed on system Y at 30, 40, 50, and 60 min after removal, following routine shaking at 40 rpm for the first 20 min. Environmental records documented ambient temperatures of 20-25 °C; specimen temperatures were not measured. ALP and six comparison enzymes were measured in technical duplicate; 30-min ALP means were submitted for the first 2026 EQA round. Relative bias was calculated against method- or peer-group consensus targets, with materials as independent statistical units. Mean ALP bias (± standard deviation) changed from -6.7% ± 1.1% at 30 min to -1.7% ± 1.0% at 50 min and -1.3% ± 1.2% at 60 min. The exploratory 30-to-50-min comparison showed less negative bias in all five materials: mean paired change, 4.9 percentage points (t-based 95% confidence interval, 4.3-5.6; paired t-test: t = 21.248, df = 4, p < 0.001; two-sided exact sign test: p = 0.063). Comparison enzymes provided descriptive context without formal between-enzyme testing. Two subsequent EQA rounds yielded mean biases of 1.6% ± 1.2% and -0.2% ± 0.7%; different materials precluded paired comparisons or intervention validation. Longer elapsed time was associated with reduced negative ALP bias in these previously frozen materials. No matched refrigerated aliquots were tested, and commutability with patient specimens was not assessed. Measurement order remained confounded with elapsed time. Findings do not establish the original deviation's cause, improved analytical trueness, or an optimal rewarming interval. They support investigation of material-specific EQA handling; extrapolation to patient specimens requires separate evidence.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Alkaline PhosphataseSpecimen HandlingQuality ControlTime Factors

Substanzen

Alkaline Phosphatase

Autor:innen und Institutionen

  • Department of Laboratory Medicine, Chuxiong Yi Autonomous Prefecture People's Hospital.
  • Department of Plastic and Burn Surgery, West China Hospital, Sichuan University.
  • Department of Laboratory Medicine, West China Hospital, Sichuan University; wuhanzhum@wchscu.cn.
  • Department of Laboratory Medicine, Chuxiong Yi Autonomous Prefecture People's Hospital; 18987838289@126.com.
  • Department of Laboratory Medicine, Chuxiong Yi Autonomous Prefecture People's Hospital; 289720630@qq.com.

Zitieren

APA

Xie, D., Wen, X., Chen, B., Li, Y., Zhu, M., Zhou, W., & Liu, J. (2026). Prolonged Rewarming Was Associated with Reduced Alkaline Phosphatase Bias in External Quality Assessment Materials: A Laboratory Study. Journal of Visualized Experiments. https://doi.org/10.3791/75749

Vancouver

Xie D, Wen X, Chen B, Li Y, Zhu M, Zhou W, et al. Prolonged Rewarming Was Associated with Reduced Alkaline Phosphatase Bias in External Quality Assessment Materials: A Laboratory Study. Journal of Visualized Experiments. 2026. doi:10.3791/75749

BibTeX
@article{Xie2026,
  author = {Xie, Dayun and Wen, Xue and Chen, Baoxiu and Li, Yali and Zhu, Man and Zhou, Wenbiao and Liu, Jiao},
  title = {Prolonged Rewarming Was Associated with Reduced Alkaline Phosphatase Bias in External Quality Assessment Materials: A Laboratory Study},
  journal = {Journal of Visualized Experiments},
  year = {2026},
  number = {237},
  doi = {10.3791/75749}
}
Kostenlos starten →