Comparison between Different Visual Acuity Tests and Validation of a Digital Device
Vision · 23.09.2024 · 8 (3) · 57
DOI: 10.3390/vision8030057 ↗Kurz gefasst
Fachartikel in der Zeitschrift Vision, online erschienen am 23.09.2024. Verfasst von Blanca Montori und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmacology and Physiology, University of Zaragoza, Domingo Miral, s/n., 50009 Zaragoza, Spain. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Institute of Health Research Carlos III, Spanish Government through the Institute of Health Research Carlos III, CDTI.
Abstract (Originalsprache)
Purpose
To compare different visual acuity (VA) tests (printed and digital, symbols and letters) and to validate a new device for VA testing called DIVE (Devices for an Integral Visual Examination).
Methods
VA was tested in a wide spectrum of adult people with printed tests (ETDRS and LEA Symbols) and with two implemented tests in DIVE (HOTV and DIVE Symbols). We measured agreement between the different VA tests using the intraclass correlation coefficient and Bland–Altman method. In addition, we measured the repeatability of all tests.
Results
Right eyes from 51 adult participants were included in the study. Correlation between tests was high (ICC from 0.95 to 0.97). Bland–Altman analysis showed good agreement among the different tests, with differences within reasonable clinical limits. However, slightly better VA values were obtained with DIVE HOTV and ETDRS, followed by LEA and DIVE Symbols. ETDRS had the best repeatability.
Conclusion
The four evaluated VA tests provide comparable outcomes. In an adult sample, letter optotypes obtained better VA values than symbol optotypes. DIVE VA tests are reliable and well-correlated with printed VA tests.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Pharmacology and Physiology, University of Zaragoza, Domingo Miral, s/n., 50009 Zaragoza, Spain.
- Aragón Health Research Institute (IIS Aragón), 50009 Zaragoza, Spain.
- Ophthalmology Department, Miguel Servet University Hospital, 50009 Zaragoza, Spain.
- Dive Medical S.L., Edificio Ceminem Spinup, Calle Mariano Esquillor Gómez, 50018 Zaragoza, Spain.
ORCID-Profile: Maria Vilella, Xian Pan, Marta Lacort
Förderung
- Institute of Health Research Carlos III
- Spanish Government through the Institute of Health Research Carlos III
- CDTI
- Seal of Excellence of the Horizon Europe EIC Accelerator
- Ministerio de Ciencia e Innovación and the Agencia Estatal de Investigación of Spain
Zitieren
APA
Montori, B., Pérez Roche, T., Vilella, M., López, E., Alejandre, A., Pan, X., Ortín, M., Lacort, M., & Pueyo, V. (2024). Comparison between Different Visual Acuity Tests and Validation of a Digital Device. Vision, 8(3), 57. https://doi.org/10.3390/vision8030057
Vancouver
Montori B, Pérez Roche T, Vilella M, López E, Alejandre A, Pan X, et al. Comparison between Different Visual Acuity Tests and Validation of a Digital Device. Vision. 2024;8(3):57. doi:10.3390/vision8030057
BibTeX
@article{Montori2024,
author = {Montori, Blanca and Pérez Roche, Teresa and Vilella, Maria and López, Estela and Alejandre, Adrián and Pan, Xian and Ortín, Marta and Lacort, Marta and Pueyo, Victoria},
title = {Comparison between Different Visual Acuity Tests and Validation of a Digital Device},
journal = {Vision},
year = {2024},
volume = {8},
number = {3},
pages = {57},
doi = {10.3390/vision8030057}
}
EUVIMED