Modifying the trajectory of epicardial leads can substantially reduce MRI ‐induced RF heating in pediatric patients with a cardiac implantable electronic device at 1.5T

Fuchang Jiang iD, Bhumi Bhusal iD, Bach Nguyen iD, Michael Monge, Gregory Webster, Daniel Kim iD, Giorgio Bonmassar iD, Andrada R Popsecu, Laleh Golestanirad

Magnetic Resonance in Medicine · 2023

DOI: 10.1002/mrm.29776 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2023. Verfasst von Fuchang Jiang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, Illinois, USA. Hauptthemen laut MEDLINE: Heating, Prostheses and Implants. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIH HHS, NIBIB NIH HHS, NHLBI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

After epicardial cardiac implantable electronic devices are implanted in pediatric patients, they become ineligible to receive MRI exams due to an elevated risk of RF heating. We investigated whether simple modifications in the trajectories of epicardial leads could substantially and reliably reduce RF heating during MRI at 1.5 T, with benefits extending to abandoned leads.

Methods

Electromagnetic simulations were performed to assess RF heating of two common 35‐cm epicardial lead trajectories exhibiting different degrees of coupling with MRI incident electric fields. Experiments in anthropomorphic phantoms implanted with commercial cardiac implantable electronic devices confirmed the findings. Both electromagnetic simulations and experimental measurements were performed using head‐first and feet‐first positioning and various landmarks. Transfer function approach was used to assess the performance of suggested modifications in realistic body models.

Results

Simulations (head‐first, chest landmark) of a 35‐cm epicardial lead with a trajectory where the excess length of the lead was looped and placed on the inferior surface of the heart showed an 87‐fold reduction in the 0.1 g–averaged specific absorption rate compared with the lead where the excess length was looped on the anterior surface. Repeated experiments with a commercial epicardial device confirmed this. For fully implanted systems following low–specific absorption rate trajectories, there was a 16‐fold reduction in the average temperature rise and a 28‐fold reduction for abandoned leads. The transfer function method predicted a 7‐fold reduction in the RF heating in 336 realistic scenarios.

Conclusion

Surgical modification of epicardial lead trajectory can substantially reduce RF heating at 1.5 T, with benefits extending to abandoned leads.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HeatingProstheses and ImplantsHumansChildHeartTemperatureMagnetic Resonance ImagingPhantoms, ImagingRadio WavesHot Temperature

Stichwörter der Autor:innen

RF heatingcardiovascular implantable electronic devicesepicardial leadssafety

Autor:innen und Institutionen

  • Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, Illinois, USA.
  • Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA.
  • Division of Cardiovascular-Thoracic Surgery, Ann & Robert H. Lurie Children's Hospital of Chicago, Chicago, Illinois, USA.
  • Division of Cardiology, Ann and Robert H. Lurie Children's Hospital of Chicago, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
  • A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Division of Medical Imaging, Ann & Robert H. Lurie Children's Hospital of Chicago, Chicago, Illinois, USA.

ORCID-Profile: Fuchang Jiang, Bhumi Bhusal, Bach Nguyen, Daniel Kim, Giorgio Bonmassar

Förderung

  • NIH HHS
  • NIBIB NIH HHS
  • NHLBI NIH HHS
  • National Institutes of Health

Zitieren

APA

Jiang, F., Bhusal, B., Nguyen, B., Monge, M., Webster, G., Kim, D., Bonmassar, G., Popsecu, A. R., & Golestanirad, L. (2023). Modifying the trajectory of epicardial leads can substantially reduce MRI ‐induced RF heating in pediatric patients with a cardiac implantable electronic device at 1.5T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29776

Vancouver

Jiang F, Bhusal B, Nguyen B, Monge M, Webster G, Kim D, et al. Modifying the trajectory of epicardial leads can substantially reduce MRI ‐induced RF heating in pediatric patients with a cardiac implantable electronic device at 1.5T. Magnetic Resonance in Medicine. 2023. doi:10.1002/mrm.29776

BibTeX
@article{Jiang2023,
  author = {Jiang, Fuchang and Bhusal, Bhumi and Nguyen, Bach and Monge, Michael and Webster, Gregory and Kim, Daniel and Bonmassar, Giorgio and Popsecu, Andrada R and Golestanirad, Laleh},
  title = {Modifying the trajectory of epicardial leads can substantially reduce MRI ‐induced RF heating in pediatric patients with a cardiac implantable electronic device at 1.5T},
  journal = {Magnetic Resonance in Medicine},
  year = {2023},
  doi = {10.1002/mrm.29776}
}