Overestimation of the Apparent Diffusion Coefficient in Diffusion-Weighted Imaging Due to Residual Fat Signal and Out-of-Phase Conditions
Tomography · 2026
DOI: 10.3390/tomography12010011 ↗Kurz gefasst
Fachartikel in der Zeitschrift Tomography, 2026. Verfasst von Maher Dhanani und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiology, Uniklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, 91052 Erlangen, Germany. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Adipose Tissue, Breast. Die Arbeit nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bavarian Academic Center for Central, Eastern, and Southeastern Europe and the German Research Foundation.
Abstract (Originalsprache)
Background/Objectives: Diffusion-weighted imaging (DWI) is a magnetic resonance technique used to map the apparent diffusion coefficient (ADC) of water in human tissue. ADC assessment plays a central role in clinical diagnostics, as malignant tissues typically exhibit reduced water mobility and, thus, lower ADC values. Accurately measuring the ADC requires effective fat suppression to prevent contamination from the residual fat signal, which is commonly believed to cause ADC underestimation. This study aimed to demonstrate that ADC overestimation may occur as well.
Methods
Our theoretical analysis shows that out-of-phase conditions between fat and water signals lead to ADC overestimations. We performed demonstration experiments on fat–water phantoms and the breasts of 10 healthy female volunteers. In particular, we considered three out-of-phase conditions: First and second, short-time inversion recovery (STIR) fat suppression with incorrect inversion time and incorrect flip angle, respectively. Third, phase differences due to spectral fat saturation. The ADC values were assessed in regions of interest (ROIs) that included both water and residual fat signals.
Results
In the phantoms and the volunteer data, ROIs containing both fat and water signals consistently exhibited lower ADC values under in-phase conditions and higher ADC values under out-of-phase conditions.
Conclusions
We demonstrated that out-of-phase conditions can result in ADC overestimation in the presence of residual fat signals, potentially resulting in false-negative classifications where malignant lesions are misinterpreted as benign due to an elevated ADC. Out-of-phase fat and water signals might also reduce lesion conspicuity in high b-value images, potentially masking clinically relevant findings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Radiology, Uniklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, 91052 Erlangen, Germany.
- Department of Medical Physics in Radiology, German Cancer Research Center, 69120 Heidelberg, Germany.
ORCID-Profile: Maher Dhanani, Dominika Skwierawska, Sabine Ohlmeyer, Frederik Bernd Laun
Förderung
- Bavarian Academic Center for Central, Eastern, and Southeastern Europe and the German Research Foundation
Zitieren
APA
Dhanani, M., Skwierawska, D., Kuder, T. A., Ohlmeyer, S., Uder, M., Bickelhaupt, S., & Laun, F. B. (2026). Overestimation of the Apparent Diffusion Coefficient in Diffusion-Weighted Imaging Due to Residual Fat Signal and Out-of-Phase Conditions. Tomography. https://doi.org/10.3390/tomography12010011
Vancouver
Dhanani M, Skwierawska D, Kuder TA, Ohlmeyer S, Uder M, Bickelhaupt S, et al. Overestimation of the Apparent Diffusion Coefficient in Diffusion-Weighted Imaging Due to Residual Fat Signal and Out-of-Phase Conditions. Tomography. 2026. doi:10.3390/tomography12010011
BibTeX
@article{Dhanani2026,
author = {Dhanani, Maher and Skwierawska, Dominika and Kuder, Tristan Anselm and Ohlmeyer, Sabine and Uder, Michael and Bickelhaupt, Sebastian and Laun, Frederik Bernd},
title = {Overestimation of the Apparent Diffusion Coefficient in Diffusion-Weighted Imaging Due to Residual Fat Signal and Out-of-Phase Conditions},
journal = {Tomography},
year = {2026},
doi = {10.3390/tomography12010011}
}