In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement

Elena V. Khoroshun, Vladimir A. Kozlov, Igor V. Ivanov, Kuvat T. Momynaliev

Sensors · 19.11.2026 · 26 (22) · 7297

DOI: 10.3390/s26227297 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 19.11.2026. Verfasst von Elena V. Khoroshun und 3 weiteren Autor:innen. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Continuous glucose monitoring (CGM) sensors are evaluated almost exclusively in clinical studies. We investigated the dynamic response of one factory-calibrated CGM system on a programmable in vitro flow platform, with glucose measured independently by a SUPER GL2 analyzer as comparator. Twelve sensors were studied: four during verification of the measurement procedure (Stage 1) and eight tested on two consecutive days (Stage 2) under a 90 → 270 → 90 mg/dL profile. Amplitude transfer (KA) and apparent rising- and falling-phase slope transfer (Kup, Kdown) were calculated without lag correction. Stage 1 results were exploratory owing to an unexplained CGM displacement of approximately 35 min. The delivered profile differed from the programmed one in every Stage 2 run (upper plateaus 254–268 vs. 270 mg/dL). Mean KA was 0.731 and 0.731, Kup 0.749 and 0.746, and Kdown 0.729 and 0.719 on the two days; between sensors, KA ranged from 0.61–0.84. Near-zero mean paired changes reflected opposite group shifts (KA +0.026 and −0.025). Descriptive ICC(A,1) values of 0.919–0.952 (seven or eight pairs from two shared runs) indicate preserved between-sensor differences rather than established reproducibility. Bias (≈−50 mg/dL) and MARD (31–32%) apply only to these in vitro conditions; the findings do not establish clinical performance or long-term stability.

Zitieren

APA

Khoroshun, E. V., Kozlov, V. A., Ivanov, I. V., & Momynaliev, K. T. (2026). In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement. Sensors, 26(22), 7297. https://doi.org/10.3390/s26227297

Vancouver

Khoroshun EV, Kozlov VA, Ivanov IV, Momynaliev KT. In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement. Sensors. 2026;26(22):7297. doi:10.3390/s26227297

BibTeX
@article{Khoroshun2026,
  author = {Khoroshun, Elena V. and Kozlov, Vladimir A. and Ivanov, Igor V. and Momynaliev, Kuvat T.},
  title = {In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {22},
  pages = {7297},
  doi = {10.3390/s26227297}
}
Kostenlos starten →