Closed-Loop Frequency Tracking of a Surface-Micromachined Thermal Sensor for Hydrogen Detection
Sensors · 10.11.2026 · 26 (22) · 7099
DOI: 10.3390/s26227099 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 10.11.2026. Verfasst von Philipp Raimann und 3 weiteren Autor:innen. Die Arbeit nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Private Stiftung Ewald Marquardt, German Federal Ministry of Education and Research.
Abstract (Originalsprache)
Hydrogen detection with high sensitivity and fast response is essential for the safe operation of hydrogen-based technologies. This work presents a surface-micromachined thermal MEMS sensor that exploits the thermophysical properties of the surrounding gas for hydrogen detection. The sensor comprises a honeycomb-shaped n-doped polysilicon heater suspended on an 80 nm-thick perforated membrane above a 1 µm gas gap, providing high thermal isolation and low thermal mass. A lumped-element model was introduced to describe the static and dynamic behavior of the sensor. The miniaturization enables dynamic operation in the kilohertz range using the 3ω method. A closed-loop phase-locked approach was implemented to continuously adjust the excitation frequency to a defined phase of the thermal response, using the resulting operating frequency as the measurement parameter. A noise-equivalent 3σ hydrogen detection limit of approximately 6 ppm was determined from the Allan deviation at an integration time of 20 s. However, the sensor response exhibits a pronounced pressure dependence due to rarefaction effects in the micrometer-scale gas gap, which modify the effective thermal conductivity and thus the dynamic thermal response. These results demonstrate the potential of high-bandwidth thermal MEMS sensors for sensitive hydrogen detection while highlighting the need to account for pressure-dependent heat transfer at microscale dimensions.
Förderung
- Private Stiftung Ewald Marquardt
- German Federal Ministry of Education and Research
Zitieren
APA
Raimann, P., Hedrich, F., Billat, S., & Dehé, A. (2026). Closed-Loop Frequency Tracking of a Surface-Micromachined Thermal Sensor for Hydrogen Detection. Sensors, 26(22), 7099. https://doi.org/10.3390/s26227099
Vancouver
Raimann P, Hedrich F, Billat S, Dehé A. Closed-Loop Frequency Tracking of a Surface-Micromachined Thermal Sensor for Hydrogen Detection. Sensors. 2026;26(22):7099. doi:10.3390/s26227099
BibTeX
@article{Raimann2026,
author = {Raimann, Philipp and Hedrich, Frank and Billat, Sophie and Dehé, Alfons},
title = {Closed-Loop Frequency Tracking of a Surface-Micromachined Thermal Sensor for Hydrogen Detection},
journal = {Sensors},
year = {2026},
volume = {26},
number = {22},
pages = {7099},
doi = {10.3390/s26227099}
}