EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Configuration-Specific Temporal Alignment of a Repurposed Wearable Fuel-Cell Alcohol Sensor with an MQ-3 Timing Proxy in a Laboratory Static-Headspace Chamber

Hasan Mirzakhaninafchi iD, Glen Rains iD, Wesley Porter iD, Canicius Mwitta iD, Denis Kiobia, Shekhar Thapa iD, Peter Ngimbwa, Christopher Ranger iD

Sensors · 03.11.2026 · 26 (21) · 6949

DOI: 10.3390/s26216949 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 03.11.2026. Verfasst von Hasan Mirzakhaninafchi und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Entomology, University of Georgia, Tifton, GA 31793, USA. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Ethanol, Alcohols, Bioelectric Energy Sources. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute of Food and Agriculture, USDA National Institute of Food and Agriculture, Crop Research Initiative.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Ethanol emitted from stressed woody tissues is a host-associated cue for ambrosia beetles, motivating investigation of portable ethanol-responsive sensors for future tree monitoring. This laboratory study evaluated temporal alignment between two asynchronously logged volatile organic compound (VOC)-responsive sensors during a denatured-alcohol chamber challenge. We selected the BACtrack Skyn wearable fuel-cell alcohol sensor for its compact form and timestamped output as a candidate for future tree-headspace measurements, and evaluated one device with an MQ-3 metal-oxide timing proxy and environmental logging in a closed static-headspace chamber. The dataset comprised five alcohol-free control runs (B0–B4) and three technical 1 μL denatured-alcohol challenge runs (EtOH1–EtOH3). Signals were processed on a common 1 min grid using a 10–15 min analytical baseline and a 15–45 min response-characterization window. MQ-3 peaks occurred at 17.0–17.3 min from the run origin (T0), whereas Skyn device-output peaks occurred at 18.8–19.2 min; peak-time offsets were 1.5–2.2 min. First-difference analysis selected a 2 min alignment setting on the primary 1 min grid in all three challenge runs; a supplementary 20 s analysis gave run-specific optima of 80–160 s. Across the three challenge runs, run-specific Pearson correlations ranged from 0.686 to 0.751 before alignment and from 0.861 to 0.930 after alignment; each coefficient was calculated from the paired 1 min observations in the 15–45 min response window. Normalized root-mean-square error (RMSE) ranged from 0.208 to 0.272 before alignment and from 0.121 to 0.230 after alignment, whereas rank correlation did not improve consistently. Alcohol-free controls remained below challenge-run peak magnitudes, although MQ-3 traces included drift and positive excursions. These within-run comparisons describe alignment in the tested configuration, not intrinsic sensor latency, concentration calibration, or validated tree-monitoring performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Wearable Electronic DevicesEthanolAlcoholsBioelectric Energy SourcesBiosensing TechniquesAnimalsVolatile Organic Compounds

Stichwörter der Autor:innen

BACtrack SkynMQ-3denatured alcoholethanol sensingmetal-oxide sensorstatic headspacetemporal alignmenttree-monitoring developmentwearable fuel-cell sensor

Substanzen

EthanolVolatile Organic CompoundsAlcohols

Autor:innen und Institutionen

  • Department of Entomology, University of Georgia, Tifton, GA 31793, USA.
  • Department of Crop and Soil Sciences, University of Georgia, Tifton, GA 31793, USA.
  • College of Engineering, University of Georgia, Tifton, GA 31793, USA.
  • Application Technology Research Unit, United States Department of Agriculture, Agricultural Research Service, 1680 Madison Avenue, Wooster, OH 44691, USA.

ORCID-Profile: Hasan Mirzakhaninafchi, Glen Rains, Wesley Porter, Canicius Mwitta, Shekhar Thapa, Christopher Ranger

Förderung

  • National Institute of Food and Agriculture
  • USDA National Institute of Food and Agriculture
  • Crop Research Initiative

Zitieren

APA

Mirzakhaninafchi, H., Rains, G., Porter, W., Mwitta, C., Kiobia, D., Thapa, S., Ngimbwa, P., & Ranger, C. (2026). Configuration-Specific Temporal Alignment of a Repurposed Wearable Fuel-Cell Alcohol Sensor with an MQ-3 Timing Proxy in a Laboratory Static-Headspace Chamber. Sensors, 26(21), 6949. https://doi.org/10.3390/s26216949

Vancouver

Mirzakhaninafchi H, Rains G, Porter W, Mwitta C, Kiobia D, Thapa S, et al. Configuration-Specific Temporal Alignment of a Repurposed Wearable Fuel-Cell Alcohol Sensor with an MQ-3 Timing Proxy in a Laboratory Static-Headspace Chamber. Sensors. 2026;26(21):6949. doi:10.3390/s26216949

BibTeX
@article{Mirzakhaninafchi2026,
  author = {Mirzakhaninafchi, Hasan and Rains, Glen and Porter, Wesley and Mwitta, Canicius and Kiobia, Denis and Thapa, Shekhar and Ngimbwa, Peter and Ranger, Christopher},
  title = {Configuration-Specific Temporal Alignment of a Repurposed Wearable Fuel-Cell Alcohol Sensor with an MQ-3 Timing Proxy in a Laboratory Static-Headspace Chamber},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {21},
  pages = {6949},
  doi = {10.3390/s26216949}
}