Design and Bench Evaluation of a Sensor-Integrated Prosthetic-Leg Prototype
Sensors · 2026
DOI: 10.3390/s26196341 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2026. Verfasst von Turar Seitkassymov und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Robotics and Technical Means of Automation, Satbayev University, Almaty 050013, Kazakhstan. Hauptthemen laut MEDLINE: Artificial Limbs, Prosthesis Design, Leg, Biosensing Techniques. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Science and Higher Education of the Republic of Kazakhstan, Science Committee of the Ministry of Science and Higher Education of the Republic of Kazakhstan.
Abstract (Originalsprache)
This study presents the design and preliminary bench evaluation of a sensor-integrated prosthetic-leg prototype. The platform combines a multi-material PAHT-CF/TPU 95A foot, articulated knee and ankle mechanisms, two surface electromyography channels, an MPU6050 inertial measurement unit, an Arduino Uno controller, and MG995 servomotors. EMG threshold crossings select discrete knee–servomotor commands, while an accelerometer-derived inclination estimate selects between two ankle–servomotor commands. The evaluation comprised exploratory finite-element analyses, separate static compression tests, and qualitative demonstrations of the sensor-to-actuator pathways. The foot model predicted a maximum displacement of 25.96 mm under 400 N, whereas a separate compression test recorded approximately 7 mm at 1000 N; these values do not constitute quantitative model validation because the loading, constraints, and displacement measures differed. The shank model produced a maximum equivalent stress of 23.89 MPa, and a separate shank–knee assembly test reached approximately 1.85 kN. The foot–ankle connection produced a stress estimate of 203.6 MPa against an assumed strength value of 60 MPa, identifying a critical design limitation. Functional observations demonstrated sensor-triggered command execution, but joint tracking accuracy, response time, actuator output under load, and control repeatability were not quantified. The prototype therefore provides an initial laboratory platform for investigating mechanical and sensing integration; structural redesign and quantitative bench testing are required before any load-bearing or clinical use can be considered.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Robotics and Technical Means of Automation, Satbayev University, Almaty 050013, Kazakhstan.
- Department of Aerospace and Electronic Engineering, Institute of Telecommunications and Automation, Almaty University of Power Engineering and Telecommunications, Almaty 050013, Kazakhstan.
- Department of Science and Innovations, Mukhametzhan Tynyshbayev ALT University, Almaty 050013, Kazakhstan.
ORCID-Profile: Yerkebulan Nurgizat, Nursultan Zhetenbayev, Aidos Sultan
Förderung
- Ministry of Science and Higher Education of the Republic of Kazakhstan
- Science Committee of the Ministry of Science and Higher Education of the Republic of Kazakhstan
Zitieren
APA
Seitkassymov, T., Nurgizat, Y., Zhetenbayev, N., Sultan, A., Sergazin, G., & Ozhikenov, K. (2026). Design and Bench Evaluation of a Sensor-Integrated Prosthetic-Leg Prototype. Sensors. https://doi.org/10.3390/s26196341
Vancouver
Seitkassymov T, Nurgizat Y, Zhetenbayev N, Sultan A, Sergazin G, Ozhikenov K. Design and Bench Evaluation of a Sensor-Integrated Prosthetic-Leg Prototype. Sensors. 2026. doi:10.3390/s26196341
BibTeX
@article{Seitkassymov2026,
author = {Seitkassymov, Turar and Nurgizat, Yerkebulan and Zhetenbayev, Nursultan and Sultan, Aidos and Sergazin, Gani and Ozhikenov, Kassymbek},
title = {Design and Bench Evaluation of a Sensor-Integrated Prosthetic-Leg Prototype},
journal = {Sensors},
year = {2026},
doi = {10.3390/s26196341}
}