Design and Optimization of an Inductive-Stub-Coupled CSRR for Non-Invasive Glucose Sensing

Zaid A Abdul Hassain, Malik J Farhan iD, Taha A Elwi, Iulia Andreea Mocanu iD

Sensors · 2025

DOI: 10.3390/s25247592 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Zaid A Abdul Hassain und 3 weiteren Autor:innen. Beteiligte Einrichtung: Electrical Engineering Department, Mustansiriyah University, Baghdad 1004, Iraq. Hauptthemen laut MEDLINE: Blood Glucose, Biosensing Techniques, Blood Glucose Self-Monitoring, Glucose. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents a high-sensitivity microwave sensor based on a modified Complementary Split Ring Resonator (CSRR) architecture, integrated with inductive stubs, for non-invasive blood glucose monitoring. The proposed sensor is designed to enhance the electric field localization and coupling efficiency by introducing inductive elements that strengthen the perturbation effect caused by glucose concentration changes in the blood. Numerical simulations were conducted using a multilayer finger model to evaluate the sensor’s performance under various glucose levels ranging from 0 to 500 mg/dL. The modified sensor exhibits dual-resonance characteristics and outperforms the conventional CSRR in both frequency and amplitude sensitivity. At an optimized stub gap of 2 mm, which effectively minimizes the capacitive coupling effect of the transmission line and thereby improves the quality factor, the sensor achieves a frequency shift sensitivity of 0.086 MHz/mg/dL and an amplitude sensitivity of 0.02 dB/mg/dL, compared to 0.032 MHz/mg/dL and 0.0116 dB/mg/dL observed in the standard CSRR structure. This confirms a significant enhancement in sensing performance and field confinement due to the optimized inductive loading. These results represent significant enhancements of approximately 168% and 72%, respectively. With its compact design, increased sensitivity, and potential for wearable implementation, the proposed sensor offers a promising platform for continuous, real-time, and non-invasive glucose monitoring in biomedical applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood GlucoseBiosensing TechniquesBlood Glucose Self-MonitoringGlucoseHumansMicrowavesEquipment DesignWearable Electronic Devices

Stichwörter der Autor:innen

CSRRglucose levelsnon-invasive glucose measurementresonator

Substanzen

Blood GlucoseGlucose

Autor:innen und Institutionen

  • Electrical Engineering Department, Mustansiriyah University, Baghdad 1004, Iraq.
  • Department of Automation and Artificial Intelligence Engineering, College of Information Engineering, Al-Nahrain University, Baghdad 10001, Iraq.
  • Telecommunications Department, National University of Science and Technology Politehnica Bucharest, 060042 Bucharest, Romania.

ORCID-Profile: Malik J Farhan, Iulia Andreea Mocanu

Zitieren

APA

Hassain, Z. A. A., Farhan, M. J., Elwi, T. A., & Mocanu, I. A. (2025). Design and Optimization of an Inductive-Stub-Coupled CSRR for Non-Invasive Glucose Sensing. Sensors. https://doi.org/10.3390/s25247592

Vancouver

Hassain ZAA, Farhan MJ, Elwi TA, Mocanu IA. Design and Optimization of an Inductive-Stub-Coupled CSRR for Non-Invasive Glucose Sensing. Sensors. 2025. doi:10.3390/s25247592

BibTeX
@article{Hassain2025,
  author = {Hassain, Zaid A Abdul and Farhan, Malik J and Elwi, Taha A and Mocanu, Iulia Andreea},
  title = {Design and Optimization of an Inductive-Stub-Coupled CSRR for Non-Invasive Glucose Sensing},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25247592}
}
Kostenlos starten →