Fachartikel in der Zeitschrift Advanced Science, online erschienen am 12.09.2024. Verfasst von Tamoghna Saha und 12 weiteren Autor:innen. Beteiligte Einrichtung: Aiiso Yufeng Li Family Department of Chemical and Nanoengineering, University of California San Diego, La Jolla, CA, 92093, USA. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Sweat, Blood Glucose Self-Monitoring. Die Arbeit wurde bisher 76-mal zitiert (Crossref) und nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Center of Wearable Sensors, Kuwait Foundation for the Advancement of Sciences.
Abstract (Originalsprache)
The demand for glucose monitoring devices has witnessed continuous growth from the rising diabetic population. The traditional approach of blood glucose (BG) sensor strip testing generates only intermittent glucose readings. Interstitial fluid‐based devices measure glucose dynamically, but their sensing approaches remain either minimally invasive or prone to skin irritation. Here, a sweat glucose monitoring system is presented, which completely operates under rest with no sweat stimulation and can generate real‐time BG dynamics. Osmotically driven hydrogels, capillary action with paper microfluidics, and self‐powered enzymatic biochemical sensor are used for simultaneous sweat extraction, transport, and glucose monitoring, respectively. The osmotic forces facilitate greater flux inflow and minimize sweat rate fluctuations compared to natural perspiration‐based sampling. The epidermal platform is tested on fingertip and forearm under varying physiological conditions. Personalized calibration models are developed and validated to obtain real‐time BG information from sweat. The estimated BG concentration showed a good correlation with measured BG concentration, with all values lying in the A+B region of consensus error grid (MARD = 10.56% (fingertip) and 13.17% (forearm)). Overall, the successful execution of such osmotically driven continuous BG monitoring system from passive sweat can be a useful addition to the next‐generation continuous sweat glucose monitors.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Aiiso Yufeng Li Family Department of Chemical and Nanoengineering, University of California San Diego, La Jolla, CA, 92093, USA.
- Department of Bioengineering, University of California San Diego, La Jolla, CA, 92093, USA.
- Department of Mechanical Engineering, University of California San Diego, La Jolla, CA, 92093, USA.
ORCID-Profile: Joseph Wang
Förderung
- Center of Wearable Sensors
- Kuwait Foundation for the Advancement of Sciences
Zitieren
APA
Saha, T., Khan, M. I., Sandhu, S. S., Yin, L., Earney, S., Zhang, C., Djassemi, O., Wang, Z., Han, J., Abdal, A., Srivatsa, S., Ding, S., & Wang, J. (2024). A Passive Perspiration Inspired Wearable Platform for Continuous Glucose Monitoring. Advanced Science, 11(41). https://doi.org/10.1002/advs.202405518
Vancouver
Saha T, Khan MI, Sandhu SS, Yin L, Earney S, Zhang C, et al. A Passive Perspiration Inspired Wearable Platform for Continuous Glucose Monitoring. Advanced Science. 2024;11(41). doi:10.1002/advs.202405518
BibTeX
@article{Saha2024,
author = {Saha, Tamoghna and Khan, Muhammad Inam and Sandhu, Samar Singh and Yin, Lu and Earney, Sara and Zhang, Chenyang and Djassemi, Omeed and Wang, Zongnan and Han, Jintong and Abdal, Abdulhameed and Srivatsa, Samarth and Ding, Shichao and Wang, Joseph},
title = {A Passive Perspiration Inspired Wearable Platform for Continuous Glucose Monitoring},
journal = {Advanced Science},
year = {2024},
volume = {11},
number = {41},
doi = {10.1002/advs.202405518}
}
EUVIMED